In-situ measurement of magnetic field gradient in a magnetic shield by a spin-exchange relaxation-free magnetometer |
Fang Jian-Cheng (房建成)a, Wang Tao (王涛)a, Zhang Hong (张红)b, Li Yang (李阳)a, Cai Hong-Wei (蔡洪炜)a |
a School of Instrument Science and Opto-Electronics Engineering, Science and Technology on Inertial Laboratory, Beihang University, Beijing 100191, China; b School of Instrument Science and Engineering, Southeast University, Nanjing 210096, China |
|
|
Abstract A method of measuring in-situ magnetic field gradient is proposed in this paper. The magnetic shield is widely used in the atomic magnetometer. However, there is magnetic field gradient in the magnetic shield, which would lead to additional gradient broadening. It is impossible to use an ex-situ magnetometer to measure magnetic field gradient in the region of a cell, whose length of side is several centimeters. The method demonstrated in this paper can realize the in-situ measurement of the magnetic field gradient inside the cell, which is significant for the spin relaxation study. The magnetic field gradients along the longitudinal axis of the magnetic shield are measured by a spin-exchange relaxation-free (SERF) magnetometer by adding a magnetic field modulation in the probe beam's direction. The transmissivity of the cell for the probe beam is always inhomogeneous along the pump beam direction, and the method proposed in this paper is independent of the intensity of the probe beam, which means that the method is independent of the cell's transmissivity. This feature makes the method more practical experimentally. Moreover, the AC-Stark shift can seriously degrade and affect the precision of the magnetic field gradient measurement. The AC-Stark shift is suppressed by locking the pump beam to the resonance of potassium's D1 line. Furthermore, the residual magnetic fields are measured with σ+- and σ--polarized pump beams, which can further suppress the effect of the AC-Stark shift. The method of measuring in-situ magnetic field gradient has achieved a magnetic field gradient precision of better than 30 pT/mm.
|
Received: 03 November 2014
Revised: 24 December 2014
Accepted manuscript online:
|
PACS:
|
07.55.Ge
|
(Magnetometers for magnetic field measurements)
|
|
07.55.-w
|
(Magnetic instruments and components)
|
|
41.20.Gz
|
(Magnetostatics; magnetic shielding, magnetic induction, boundary-value problems)
|
|
32.60.+i
|
(Zeeman and Stark effects)
|
|
Fund: Project supported by the National Natural Science Foundation of China (Grant Nos. 61227902, 61374210, and 61121003). |
Corresponding Authors:
Wang Tao
E-mail: wangtao@buaa.edu.cn
|
About author: 07.55.Ge; 07.55.-w; 41.20.Gz; 32.60.+i |
Cite this article:
Fang Jian-Cheng (房建成), Wang Tao (王涛), Zhang Hong (张红), Li Yang (李阳), Cai Hong-Wei (蔡洪炜) In-situ measurement of magnetic field gradient in a magnetic shield by a spin-exchange relaxation-free magnetometer 2015 Chin. Phys. B 24 060702
|
[1] |
Xia H, Baranga A B-A, Hoffman D and Romalis M V 2006 Appl. Phys. Lett. 89 211104
|
[2] |
Ledbetter M P, Savukov I M, Acosta V M, Budker D and Romalis M V 2008 Phys. Rev. A 77 033408
|
[3] |
Wyllie R, Kauer M, Wakai R T and Walker T G 2012 Opt. Lett. 37 2247
|
[4] |
Li S G, Xu Y F, Wang Z Y, Liu Y X and Lin Q 2009 Chin. Phys. Lett. 26 067805
|
[5] |
Yang A L, Yang G Q, Xu Y F and Lin Q 2014 Chin. Phys. B 23 027601
|
[6] |
Fang J C, Wang T, Zhang H, Li Y and Zou S 2014 Rev. Sci. Instrum. 85 123104
|
[7] |
Seltzer S J 2008 "Developments in Alkali-Metal Atomic Magnetometry", Ph. D. Dissertation (New Jersey: Princeton University)
|
[8] |
Sander T H, Preusser J, Mhaskar R, Kitching J, Trahms L and Knappe S 2012 Biomed. Opt. Express 3 981
|
[9] |
Wyllie R, Kauer M, Smetana G S, Wakai R T and Walker T G 2012 Phys. Med. Biol. 57 2619
|
[10] |
Kornack T W 2005 "A Test of CPT and Lorentz Symmetry Using a K-3He Co-magnetometer", Ph. D. Dissertation (New Jersey: Princeton University)
|
[11] |
Dang H B, Maloof A C and Romalis M V 2010 Appl. Phys. Lett. 97 151110
|
[12] |
Allred J C and Lyman R N 2002 Phys. Rev. Lett. 89 130801
|
[13] |
Shi R Y and Wang Y H 2013 Chin. Phys. B 22 100703
|
[14] |
Huang K K, Li N and Lu X H 2012 Chin. Phys. Lett. 29 100701
|
[15] |
Budker D and Romalis M 2007 Nat. Phys. 3 227
|
[16] |
Sumner T J, Pendlebury J M and Smith K F 1987 J. Phys. D: Appl. Phys. 20 1095
|
[17] |
Kominis I K, Kornack T W, Allred J C and Romalis M V 2003 Nature 422 596
|
[18] |
Fang J C and Qin J 2012 Rev. Sci. Instrum. 83 103104
|
[19] |
Seltzer S J and Romalis M V 2004 Appl. Phys. Lett. 85 4804
|
[20] |
Fang J C, Wang T, Quan W, Yuan H, Zhang H, Li Y and Zou S 2014 Rev. Sci. Instrum. 85 063108
|
[21] |
Bloch F 1946 Phys. Rev. 70 460
|
[22] |
Brown J M 2011 "A New Limit on Lorentz- and CPT-Violating Neutron Spin Interactions Using a K-3He Comagnetometer", Ph. D. Dissertation (New Jersey: Princeton University)
|
[23] |
Sulai I A, Wyllie R, Kauer M, Smetana G S, Wakai R T and Walker T G 2013 Opt. Lett. 38 974
|
No Suggested Reading articles found! |
|
|
Viewed |
|
|
|
Full text
|
|
|
|
|
Abstract
|
|
|
|
|
Cited |
|
|
|
|
Altmetric
|
blogs
Facebook pages
Wikipedia page
Google+ users
|
Online attention
Altmetric calculates a score based on the online attention an article receives. Each coloured thread in the circle represents a different type of online attention. The number in the centre is the Altmetric score. Social media and mainstream news media are the main sources that calculate the score. Reference managers such as Mendeley are also tracked but do not contribute to the score. Older articles often score higher because they have had more time to get noticed. To account for this, Altmetric has included the context data for other articles of a similar age.
View more on Altmetrics
|
|
|