INTERDISCIPLINARY PHYSICS AND RELATED AREAS OF SCIENCE AND TECHNOLOGY |
Prev
Next
|
|
|
Multi-objective optimization of gradient coil for benchtop magnetic resonance imaging system with high-resolution |
Wang Long-Qing (王龙庆), Wang Wei-Min (王为民) |
Institute of Quantum Electronics, School of Electronic Engineering and Computer Science, Peking University, Beijing 100871, China |
|
|
Abstract Significant high magnetic gradient field strength is essential to obtaining high-resolution images in a benchtop magnetic resonance imaging (BT-MRI) system with permanent magnet. Extending minimum wire spacing and maximum wire width of gradient coils is one of the key solutions to minimize the maximum current density so as to reduce the local heating and generate higher magnetic field gradient strength. However, maximum current density is hard to optimize together with field linearity, stored magnetic energy, and power dissipation by the traditional target field method. In this paper, a new multi-objective method is proposed to optimize the maximum current density, field linearity, stored magnetic energy, and power dissipation in MRI gradient coils. The simulation and experimental results show that the minimum wire spacings are improved by 159% and 62% for the transverse and longitudinal gradient coil respectively. The maximum wire width increases from 0.5 mm to 1.5 mm. Maximum gradient field strengths of 157 mT/m and 405 mT/m for transverse and longitudinal coil are achieved, respectively. The experimental results in BT-MRI instrument demonstrate that the MRI images with in-plane resolution of 50 μm can be obtained by using the designed coils.
|
Received: 15 April 2013
Revised: 03 May 2013
Accepted manuscript online:
|
PACS:
|
87.61.-c
|
(Magnetic resonance imaging)
|
|
84.32.Hh
|
(Inductors and coils; wiring)
|
|
02.30.Zz
|
(Inverse problems)
|
|
Fund: Project supported by the Natural Science Foundation of the Ministry of Science and Technology of China (Grant No. 2011ZX05008004) and the Science Fund of the Committee of Science and Technology of Beijing, China. |
Corresponding Authors:
Wang Wei-Min
E-mail: wmw@pku.edu.cn
|
About author: 87.61.-c; 84.32.Hh; 02.30.Zz |
Cite this article:
Wang Long-Qing (王龙庆), Wang Wei-Min (王为民) Multi-objective optimization of gradient coil for benchtop magnetic resonance imaging system with high-resolution 2014 Chin. Phys. B 23 028703
|
[1] |
Choquet P, Breton E, Goetz C, Marin C and Constantinesco A 2009 Phys. Med. Biol. 54 5287
|
[2] |
Malaterre V, Metz V, Ogorka J, Gurny R, Loggia N and Mäder K 2009 J. Control. Release 133 31
|
[3] |
Metz H and Mäder K 2008 Int. J. Pharm. 364 170
|
[4] |
Hidalgo-Tobon S S 2010 Concepts Magn. Reson. A 36 223
|
[5] |
Turner R 1986 J. Phys. D: Appl. Phys. 19 147
|
[6] |
Turner R 1988 J. Phys. E: Sci. Instrum. 21 948
|
[7] |
Brideson M A, Forbes K L and Crozier S 2002 Concepts Magn. Reson. 14 9
|
[8] |
Lemdisaov R A and Ludwig R A 2005 Magn. Reson. B: Magn.Reson. Eng. 26 67
|
[9] |
Yoda K 1990 J. Appl. Phys. 67 4349
|
[10] |
Carlson J W, Derby K A, Hawryszko K C and Weideman M 1992 Magn. Reson. Med. 26 191
|
[11] |
Liu H and Truwit C L 1998 IEEE Trans. Med. Imag. 17 826
|
[12] |
Liu W T, Zu D L and Tang X 2010 Chin. Phys. B 19 018701
|
[13] |
Forbes L K and Crozier S 2001 J. Phys. D: Appl. Phys. 34 3447
|
[14] |
Forbes L K and Crozier S 2002 J. Phys. D: Appl. Phys. 35 839
|
[15] |
Forbes L K and Crozier S 2004 IEEE Trans. Magn. 40 1929
|
[16] |
Forbes L K, Brideson M A and Crozier S 2005 IEEE Trans. Magn. 41 2134
|
[17] |
Morreone T US Patent Number 5 760 582 1998
|
[18] |
Liu W T, Zu D L, Tang X and Guo H 2007 J. Phys. D: Appl. Phys. 40 4418
|
[19] |
Hu C, Kwok J and Pan W 2009 In NIPS 781
|
[20] |
Zhang R, Xu J, Fu Y Y, Li Y J, Huang K F, Zhang J and Fang J 2011 Meas. Sci. Technol. 22 125505
|
[21] |
Poole M, Weiss P, Lopez H S, Ng M and Crozier S 2010 J. Phys. D: Appl. Phys. 43 095001
|
No Suggested Reading articles found! |
|
|
Viewed |
|
|
|
Full text
|
|
|
|
|
Abstract
|
|
|
|
|
Cited |
|
|
|
|
Altmetric
|
blogs
Facebook pages
Wikipedia page
Google+ users
|
Online attention
Altmetric calculates a score based on the online attention an article receives. Each coloured thread in the circle represents a different type of online attention. The number in the centre is the Altmetric score. Social media and mainstream news media are the main sources that calculate the score. Reference managers such as Mendeley are also tracked but do not contribute to the score. Older articles often score higher because they have had more time to get noticed. To account for this, Altmetric has included the context data for other articles of a similar age.
View more on Altmetrics
|
|
|