Please wait a minute...
Chin. Phys. B, 2014, Vol. 23(2): 028703    DOI: 10.1088/1674-1056/23/2/028703
INTERDISCIPLINARY PHYSICS AND RELATED AREAS OF SCIENCE AND TECHNOLOGY Prev   Next  

Multi-objective optimization of gradient coil for benchtop magnetic resonance imaging system with high-resolution

Wang Long-Qing (王龙庆), Wang Wei-Min (王为民)
Institute of Quantum Electronics, School of Electronic Engineering and Computer Science, Peking University, Beijing 100871, China
Abstract  Significant high magnetic gradient field strength is essential to obtaining high-resolution images in a benchtop magnetic resonance imaging (BT-MRI) system with permanent magnet. Extending minimum wire spacing and maximum wire width of gradient coils is one of the key solutions to minimize the maximum current density so as to reduce the local heating and generate higher magnetic field gradient strength. However, maximum current density is hard to optimize together with field linearity, stored magnetic energy, and power dissipation by the traditional target field method. In this paper, a new multi-objective method is proposed to optimize the maximum current density, field linearity, stored magnetic energy, and power dissipation in MRI gradient coils. The simulation and experimental results show that the minimum wire spacings are improved by 159% and 62% for the transverse and longitudinal gradient coil respectively. The maximum wire width increases from 0.5 mm to 1.5 mm. Maximum gradient field strengths of 157 mT/m and 405 mT/m for transverse and longitudinal coil are achieved, respectively. The experimental results in BT-MRI instrument demonstrate that the MRI images with in-plane resolution of 50 μm can be obtained by using the designed coils.
Keywords:  magnetic resonance imaging      gradient coil      target field method      wire spacing  
Received:  15 April 2013      Revised:  03 May 2013      Accepted manuscript online: 
PACS:  87.61.-c (Magnetic resonance imaging)  
  84.32.Hh (Inductors and coils; wiring)  
  02.30.Zz (Inverse problems)  
Fund: Project supported by the Natural Science Foundation of the Ministry of Science and Technology of China (Grant No. 2011ZX05008004) and the Science Fund of the Committee of Science and Technology of Beijing, China.
Corresponding Authors:  Wang Wei-Min     E-mail:  wmw@pku.edu.cn
About author:  87.61.-c; 84.32.Hh; 02.30.Zz

Cite this article: 

Wang Long-Qing (王龙庆), Wang Wei-Min (王为民) Multi-objective optimization of gradient coil for benchtop magnetic resonance imaging system with high-resolution 2014 Chin. Phys. B 23 028703

[1] Choquet P, Breton E, Goetz C, Marin C and Constantinesco A 2009 Phys. Med. Biol. 54 5287
[2] Malaterre V, Metz V, Ogorka J, Gurny R, Loggia N and Mäder K 2009 J. Control. Release 133 31
[3] Metz H and Mäder K 2008 Int. J. Pharm. 364 170
[4] Hidalgo-Tobon S S 2010 Concepts Magn. Reson. A 36 223
[5] Turner R 1986 J. Phys. D: Appl. Phys. 19 147
[6] Turner R 1988 J. Phys. E: Sci. Instrum. 21 948
[7] Brideson M A, Forbes K L and Crozier S 2002 Concepts Magn. Reson. 14 9
[8] Lemdisaov R A and Ludwig R A 2005 Magn. Reson. B: Magn.Reson. Eng. 26 67
[9] Yoda K 1990 J. Appl. Phys. 67 4349
[10] Carlson J W, Derby K A, Hawryszko K C and Weideman M 1992 Magn. Reson. Med. 26 191
[11] Liu H and Truwit C L 1998 IEEE Trans. Med. Imag. 17 826
[12] Liu W T, Zu D L and Tang X 2010 Chin. Phys. B 19 018701
[13] Forbes L K and Crozier S 2001 J. Phys. D: Appl. Phys. 34 3447
[14] Forbes L K and Crozier S 2002 J. Phys. D: Appl. Phys. 35 839
[15] Forbes L K and Crozier S 2004 IEEE Trans. Magn. 40 1929
[16] Forbes L K, Brideson M A and Crozier S 2005 IEEE Trans. Magn. 41 2134
[17] Morreone T US Patent Number 5 760 582 1998
[18] Liu W T, Zu D L, Tang X and Guo H 2007 J. Phys. D: Appl. Phys. 40 4418
[19] Hu C, Kwok J and Pan W 2009 In NIPS 781
[20] Zhang R, Xu J, Fu Y Y, Li Y J, Huang K F, Zhang J and Fang J 2011 Meas. Sci. Technol. 22 125505
[21] Poole M, Weiss P, Lopez H S, Ng M and Crozier S 2010 J. Phys. D: Appl. Phys. 43 095001
[1] Three-dimensional clogging structures of granular spheres near hopper orifice
Jing Yang(杨敬), Dianjinfeng Gong(宫殿锦丰), Xiaoxue Wang(汪晓雪), Zhichao Wang(王志超), Jianqi Li(李建奇), Bingwen Hu(胡炳文), and Chengjie Xia(夏成杰). Chin. Phys. B, 2022, 31(1): 014501.
[2] Design of small-scale gradient coils in magnetic resonance imaging by using the topology optimization method
Hui Pan(潘辉), Feng Jia(贾峰), Zhen-Yu Liu(刘震宇), Maxim Zaitsev, Juergen Hennig, Jan G Korvink. Chin. Phys. B, 2018, 27(5): 050201.
[3] Equivalent magnetic dipole method used to design gradient coil for unilateral magnetic resonance imaging
Zheng Xu(徐征), Xiang Li(李想), Pan Guo(郭盼), Jia-Min Wu(吴嘉敏). Chin. Phys. B, 2018, 27(5): 058702.
[4] Performance study of aluminum shielded room for ultra-low-field magnetic resonance imaging based on SQUID: Simulations and experiments
Bo Li(李波), Hui Dong(董慧), Xiao-Lei Huang(黄小磊), Yang Qiu(邱阳), Quan Tao(陶泉), Jian-Ming Zhu(朱建明). Chin. Phys. B, 2018, 27(2): 020701.
[5] Multifractal analysis of white matter structural changes on 3D magnetic resonance imaging between normal aging and early Alzheimer's disease
Ni Huang-Jing (倪黄晶), Zhou Lu-Ping (周泸萍), Zeng Peng (曾彭), Huang Xiao-Lin (黄晓林), Liu Hong-Xing (刘红星), Ning Xin-Bao (宁新宝), the Alzheimer's Disease Neuroimaging Initiative. Chin. Phys. B, 2015, 24(7): 070502.
[6] Linear-fitting-based similarity coefficient map for tissue dissimilarity analysis in T2*-w magnetic resonance imaging
Yu Shao-De (余绍德), Wu Shi-Bin (伍世宾), Wang Hao-Yu (王浩宇), Wei Xin-Hua (魏新华), Chen Xin (陈鑫), Pan Wan-Long (潘万龙), Hu Jiani, Xie Yao-Qin (谢耀钦). Chin. Phys. B, 2015, 24(12): 128711.
[7] Novel magnetic vortex nanorings/nanodiscs: Synthesis and theranostic applications
Liu Xiao-Li (刘晓丽), Yang Yong (杨勇), Wu Jian-Peng (吴建鹏), Zhang Yi-Fan (张艺凡), Fan Hai-Ming (樊海明), Ding Jun (丁军). Chin. Phys. B, 2015, 24(12): 127505.
[8] Self-assembled superparamagnetic nanoparticles as MRI contrast agents–A review
Su Hong-Ying (苏红莹), Wu Chang-Qiang (吴昌强), Li Dan-Yang (李丹阳), Ai Hua (艾华). Chin. Phys. B, 2015, 24(12): 127506.
[9] Flexible reduced field of view magnetic resonance imaging based on single-shot spatiotemporally encoded technique
Li Jing (李敬), Cai Cong-Bo (蔡聪波), Chen Lin (陈林), Chen Ying (陈颖), Qu Xiao-Bo (屈小波), Cai Shu-Hui (蔡淑惠). Chin. Phys. B, 2015, 24(10): 108703.
[10] Surface modification of magnetic nanoparticles in biomedicine
Chu Xin (储鑫), Yu Jing (余靓), Hou Yang-Long (侯仰龙). Chin. Phys. B, 2015, 24(1): 014704.
[11] Formation of multifunctional Fe3O4/Au composite nanoparticles for dual-mode MR/CT imaging applications
Hu Yong (胡勇), Li Jing-Chao (李静超), Shen Ming-Wu (沈明武), Shi Xiang-Yang (史向阳). Chin. Phys. B, 2014, 23(7): 078704.
[12] Nanomagnetism:Principles, nanostructures, and biomedical applications
Yang Ce (杨策), Hou Yang-Long (侯仰龙), Gao Song (高松). Chin. Phys. B, 2014, 23(5): 057505.
[13] Parallel variable-density spiral imaging using nonlocal total variation reconstruction
Fang Sheng (方晟), Guo Hua (郭华). Chin. Phys. B, 2014, 23(5): 057401.
[14] Multifunctional magnetic nanoparticles for magnetic resonance image-guided photothermal therapy for cancer
Yue Xiu-Li (岳秀丽), Ma Fang (马放), Dai Zhi-Fei (戴志飞). Chin. Phys. B, 2014, 23(4): 044301.
[15] Magnetic nanoparticle-based cancer therapy
Yu Jing (余靓), Huang Dong-Yan (黄冬雁), Muhammad Zubair Yousaf, Hou Yang-Long (侯仰龙), Gao Song (高松). Chin. Phys. B, 2013, 22(2): 027506.
No Suggested Reading articles found!