Chin. Phys. B, 2018, Vol. 27(12): 123701    DOI: 10.1088/1674-1056/27/12/123701
ATOMIC AND MOLECULAR PHYSICS Prev   Next  

Calibration of the superconducting gravimeter based on a cold atom absolute gravimeter at NIM

Qiyu Wang(王启宇)1, Jinyang Feng(冯金扬)1, Shaokai Wang(王少凯)1, Wei Zhuang(庄伟)1, Yang Zhao(赵阳)1, Lishuang Mou(牟丽爽)2, Shuqing Wu(吴书清)1
1 National Institute of Metrology, Beijing 100029, China;
2 China Jiliang University, Hangzhou 310018, China
Abstract  

The scale factor of a superconducting gravimeter (SG) is usually calibrated by using simultaneous and co-located gravity measurements with the FG5-type absolute gravimeter (AG). In this paper, another new kind of absolute gravimeter-cold atom gravimeter (CAG) is first reported to calibrate the SG. Five-day side-by-side gravity measurements have been carried out by using our CAG (NIM-AGRb-1) to calibrate the SG (iGrav-012) located at Changping Campus of the National Institute of Metrology (NIM) of China. A weighted least-squares method is applied to determine the scale factor and the result is given as (-928.01±0.73) nm·-2·V-1 with a precision of 0.79‰. We have demonstrated that a calibration precision of 1‰ level can be achieved after 3 days of parallel observations at spring tide. The obtained calibration results are then compared with the previous calibration by FG5X-249, which shows that the calibration precision obtained by using NIM-AGRb-1 was slightly higher than FG5X-249 with the same time interval. The factors affecting the calibration precision are analyzed in the calibrations by means of different AGs. Finally, several calibration experiments for SG iGrav-012 are discussed. The final scale factor is estimated as (-927.58±0.36) nm·-2·V-1 with an accuracy of 0.39‰. Our main results demonstrate that the CAGs can be used for high-precision calibrations of SGs.

Keywords:  cold atom gravimeter      calibration      superconducting gravimeter      scale factor  
Received:  08 June 2018      Revised:  25 September 2018      Accepted manuscript online: 
PACS:  37.25.+k (Atom interferometry techniques)  
  91.10.Pp (Geodetic techniques; gravimetric measurements and instruments)  
Fund: 

Project supported by the National Key Research and Development Program of China (Grant Nos. 2016YFF0200103 and 2016YFF0200200) and the Fundamental Research Funds for National Institute of Metrology, China (Grant No. 22-AKY1608).

Corresponding Authors:  Shuqing Wu     E-mail:  wushq@nim.ac.cn

Cite this article: 

Qiyu Wang(王启宇), Jinyang Feng(冯金扬), Shaokai Wang(王少凯), Wei Zhuang(庄伟), Yang Zhao(赵阳), Lishuang Mou(牟丽爽), Shuqing Wu(吴书清) Calibration of the superconducting gravimeter based on a cold atom absolute gravimeter at NIM 2018 Chin. Phys. B 27 123701

[1] Camp M V, Viron O, PajotM étivier G, Casenave F, Watlet A, Dassargues A and Vanclooster M 2016 Geophys. Res. Lett. 43 1
[2] Boy J P and Hinderer J 2006 J. Geodyn. 41 227
[3] Rosat S, Hinderer J, Crossley D and Boy J P 2004 J. Geodyn. 38 461
[4] Hinderer J and Crossley D 2000 Surv. Geophys. 21 1
[5] Baker T F and Bos M S 2003 Geophys. J. Int. 152 468
[6] Wu S Q, Li C J, Xu J Y, Su D W, Feng J Y and Ji W X 2017 Acta Metrological Sinca 38 127 (in Chinese)
[7] Achilli V, Baldi P, Casula G, Errani M, Focardi S, Guerzoni M, Palmonari F and Raguní G 1995 Bull. Gíodésique 69 73
[8] Falk R, Harnisch M, Harnisch G, Nowak I and Richter B 2001 J. Geod. Soc. Jpn. 47 22
[9] Melchior P 1994 Phys. Earth Planet. 82 125
[10] Hinderer J, Florsch N, M? kinen J, Legros H and Faller J E 1991 Geophys. J. R. Astron. Soc. 106 491
[11] Hinderer J, Florsch N, Mäkinen J, Legros H and Faller J E 1991 Geophys. J. R. Astron. Soc. 106 491
[12] Riccardi U, Rosat S and Hinderer J 2012 Pure Appl. Geophys. 169 1343
[13] Hwang C, Kao R, Cheng C C, Huang J F, Lee C W and Sato T 2009 Journal of Geophysical Research: Solid Earth 114 B07406
[14] Imanishi Y, Higashi T and Fukuda Y 2002 Geophys. J. Int. 151 867
[15] Francis O, Niebauer T M, Sasagawa G, Klopping F and Gschwind J 1998 Geophys. Res. Lett. 25 1075
[16] Camp M V, Meurers B, Viron O D and Forbriger T 2016 J. Geodesy 90 91
[17] Peters A, Chung K Y and Chu S 2001 Metrologia 38 25
[18] Zhou M K, Duan X C, Chen L L, Luo Q, Xu Y Y and Hu Z K 2015 Chin. Phys. B 24 050401
[19] Zhou L, Xiong Z Y, Yang W, Tang B, Peng W C, Hao K, Li R B, Liu M, Wang J and Zhan M S 2011 Gen. Relat. Gravit. 43 1931
[20] Wu B, Wang Z, Cheng B, Wang Q, Xu A and Lin Q 2014 Metrologia 51 452
[21] Hu Z K, Sun B L, Duan X C, Zhou M K, Chen L L, Zhan S, Zhang Q Z and Luo J 2013 Phys. Rev. A 88 043610
[22] Michel V C, P W S D and Olivier F 2005 J. Geophys. Res. 110 B05406
[23] Wang S K, Zhao Y, Zhuang W, Li T C, Wu S Q, Feng J Y and Li C J 2018 Metrologia 55 360
[24] Francis O and Dam T V 2002 Metrologia 39 485
[25] Meurers B 2012 Int. J. Geophys. 2012 954271
[26] Sun H P, Liu Q C, Wu S Q, Chen X D and Feng J Y 2017 Chin. J. Geophys. 60 4699 (in Chinese)
[27] Tamura Y, Sato T, Fukuda Y and Higashi T 2004 J. Geodesy 78 481
[28] Angelis M d, Bertoldi A, Cacciapuoti L, Giorgini A, Lamporesi G, Prevedelli M, Saccorotti G, Sorrentino F and Tino G M 2009 Meas. Sci. Technol. 20 022001
[29] Hauth M, Freier C, Schkolnik V, Senger A, Schmidt M and Peters A 2013 Appl. Phys. B 113 49
[30] Dickerson S M, Hogan J M, Sugarbaker A, Johnson D M and Kasevich M A 2013 Phys. Rev. Lett. 111 083001
[31] Feng J, Wu S, Li C, Su D, Xu J and Yu M 2016 Seventh International Symposium on Precision Mechanical Measurements, August 8-12, 2015, Xiamen, China, p. 990310
[1] In-situ ultrasonic calibrations of pressure and temperature in a hinge-type double-stage cubic large volume press
Qingze Li(李青泽), Xiping Chen(陈喜平), Lei Xie(谢雷), Tiexin Han(韩铁鑫), Jiacheng Sun(孙嘉程), and Leiming Fang(房雷鸣). Chin. Phys. B, 2022, 31(6): 060702.
[2] Upgrade of the magnetic diagnostic system for restart of HT-6M operation
Li-Xing Chen(陈力行), Biao Shen(沈飊), Da-Long Chen(陈大龙), Zheng-Ping Luo(罗正平),Zu-Chao Zhang(张祖超), Ying Chen(陈颖), Yong Wang(王勇), and Jin-Ping Qian(钱金平). Chin. Phys. B, 2022, 31(12): 125203.
[3] Experimental demonstration of a fast calibration method for integrated photonic circuits with cascaded phase shifters
Junqin Cao(曹君勤), Zhixin Chen(陈志歆), Yaxin Wang(王亚新), Tianfeng Feng(冯田峰), Zhihao Li(李志浩), Zeyu Xing(邢泽宇), Huashan Li(李华山), and Xiaoqi Zhou(周晓祺). Chin. Phys. B, 2022, 31(11): 114204.
[4] An extended smart driver model considering electronic throttle angle changes with memory
Congzhi Wu(武聪智), Hongxia Ge(葛红霞), and Rongjun Cheng(程荣军). Chin. Phys. B, 2022, 31(1): 010504.
[5] A pressure-calibration method of wavelength modulation spectroscopy in sealed microbial growth environment
Kun-Yang Wang(王坤阳), Jie Shao(邵杰), Li-Gang Shao(邵李刚), Jia-Jin Chen(陈家金), Gui-Shi Wang(王贵师), Kun Liu(刘琨), and Xiao-Ming Gao(高晓明). Chin. Phys. B, 2021, 30(5): 054203.
[6] A fast and precise three-dimensional measurement system based on multiple parallel line lasers
Yao Wang(王尧) and Bin Lin(林斌). Chin. Phys. B, 2021, 30(2): 024201.
[7] A novel multifunctional electronic calibration kit integrated by MEMS SPDT switches
Shan-Shan Wang(王姗姗), Qian-Nan Wu(吴倩楠), Yue-Sheng Gao(高跃升), Jian-Gang Yu(余建刚), Qian-Long Cao(曹钎龙), Lu-Lu Han(韩路路), and Meng-Wei Li(李孟委). Chin. Phys. B, 2021, 30(11): 118501.
[8] Raman scattering from highly-stressed anvil diamond
Shan Liu(刘珊), Qiqi Tang(唐琦琪), Binbin Wu(吴彬彬), Feng Zhang(张峰), Jingyi Liu(刘静仪), Chunmei Fan(范春梅), and Li Lei(雷力). Chin. Phys. B, 2021, 30(1): 016301.
[9] Analysis of extreme ultraviolet spectral profiles of laser-produced Cr plasmas
L Wu(吴磊), M G Su(苏茂根), Q Min(敏琦), S Q Cao(曹世权), S Q He(何思奇), D X Sun(孙对兄), C Z Dong(董晨钟). Chin. Phys. B, 2019, 28(7): 075201.
[10] Digitally calibrated broadband dual-comb gases absorption spectral measurements
Xinyi Chen(陈馨怡), Weipeng Zhang(张伟鹏), Haoyun Wei(尉昊赟), Yan Li(李岩). Chin. Phys. B, 2019, 28(6): 060703.
[11] Calibration and data restoration of light field modulated imaging spectrometer
Li-Juan Su(苏丽娟), Qiang-Qiang Yan(严强强), Yan Yuan(袁艳), Shi-Feng Wang(王世丰), Yu-Jian Liu(刘宇健). Chin. Phys. B, 2018, 27(8): 080702.
[12] Error analysis and optimal design of polarization calibration unit for solar telescope
Jun-Feng Hou(侯俊峰), Dong-Guang Wang(王东光), Yuan-Yong Deng(邓元勇), Ying-Zi Sun(孙英姿), Zhi-Yong Zhang(张志勇). Chin. Phys. B, 2017, 26(8): 089501.
[13] Detailed calibration of the PI-LCX: 1300 high performance single photon counting hard x-ray CCD camera
Wei Hong(洪伟), Xian-Lun Wen(温贤伦), Lai Wei(魏来), Bin Zhu(朱斌), Yu-Chi Wu(吴玉迟), Ke-Gong Dong(董克攻), Chun-Ye Jiao(焦春晔), Bo Wu(伍波), Ying-Ling He(何颖玲), Fa-Qiang Zhang(张发强), Wei-Min Zhou(周维民), Yu-Qiu Gu(谷渝秋). Chin. Phys. B, 2017, 26(2): 025204.
[14] Calibration chain design based on integrating sphere transfer radiometer for SI-traceable on-orbit spectral radiometric calibration and its uncertainty analysis
Wei-Ning Zhao(赵维宁), Wei Fang(方伟), Li-Wei Sun(孙立微), Li-Hong Cui(崔立红), Yu-Peng Wang(王玉鹏). Chin. Phys. B, 2016, 25(9): 090701.
[15] An efficient calibration method for SQUID measurement system using three orthogonal Helmholtz coils
Hua Li(李华), Shu-Lin Zhang(张树林), Chao-Xiang Zhang(张朝祥), Xiang-Yan Kong(孔祥燕), Xiao-Ming Xie(谢晓明). Chin. Phys. B, 2016, 25(6): 068501.
No Suggested Reading articles found!