Accuracy and effectiveness of self-gating signals in free-breathing three-dimensional cardiac cine magnetic resonance imaging

Shuo Li(李硕)1,3, Lei Wang(王磊)2, Yan-Chun Zhu(朱艳春)3,4, Jie Yang(杨洁)3, Yao-Qin Xie(谢耀钦)3, Nan Fu(付楠)3, Yi Wang(王乙)4, Song Gao(高嵩)1
1. Medical Imaging Physics Laboratory, Health Science Center, Peking University, Beijing 100191, China;
2. Department of Orthopedic Surgery, People's Hospital of Luohu, Shenzhen University, Shenzhen 518001, China;
3. Institute of Biomedical and Health Engineering, Shenzhen Institutes of Advanced Technology, Chinese Academy of Sciences, Shenzhen 518055, China;
4. Department of Radiology, Weill Cornell Medical College, Cornell University, New York, NY 10021, USA
Abstract  

Conventional multiple breath-hold two-dimensional (2D) balanced steady-state free precession (SSFP) presents many difficulties in cardiac cine magnetic resonance imaging (MRI). Recently, a self-gated free-breathing three-dimensional (3D) SSFP technique has been proposed as an alternative in many studies. However, the accuracy and effectiveness of self-gating signals have been barely studied before. Since self-gating signals are crucially important in image reconstruction, a systematic study of self-gating signals and comparison with external monitored signals are needed.
Previously developed self-gated free-breathing 3D SSFP techniques are used on twenty-eight healthy volunteers. Both electrocardiographic (ECG) and respiratory bellow signals are also acquired during the scan as external signals. Self-gating signal and external signal are compared by trigger and gating window. Gating window is proposed to evaluate the accuracy and effectiveness of respiratory self-gating signal. Relative deviation of the trigger and root-mean-square-deviation of the cycle duration are calculated. A two-tailed paired t-test is used to identify the difference between self-gating and external signals. A Wilcoxon signed rank test is used to identify the difference between peak and valley self-gating triggers.
The results demonstrate an excellent correlation (P=0, R>0.99) between self-gating and external triggers. Wilcoxon signed rank test shows that there is no significant difference between peak and valley self-gating triggers for both cardiac (H=0, P>0.10) and respiratory (H=0, P>0.44) motions. The difference between self-gating and externally monitored signals is not significant (two-tailed paired-sample t-test:H=0, P>0.90).
The self-gating signals could demonstrate cardiac and respiratory motion accurately and effectively as ECG and respiratory bellow. The difference between the two methods is not significant and can be explained. Furthermore, few ECG trigger errors appear in some subjects while these errors are not found in self-gating signals.

Keywords:  self-gating      free-breathing      cine MRI      effectiveness      accuracy  
Received:  19 May 2016      Revised:  17 August 2016      Accepted manuscript online: 
PACS:  87.61.-c (Magnetic resonance imaging)  
  87.57.-s (Medical imaging)  
Fund: 

Project supported by the National Natural Science Foundation of China (Grant Nos. 81501463, 61471349, 81671853, 81571669, and 61671026), the National High Technology Research and Development Program of China (Grant No. 2015AA043203), the Natural Science Foundation of Beijing, China (Grant No. 7162112), Guangdong Innovative Research Team Program of China (Grant No. 2011S013), the Natural Science Foundation of Guangdong Province, China (Grant Nos. 2014A030310360 and 2014A0202015028), the Beijing Center for Mathematics and Information Interdisciplinary Sciences, Shenzhen Fundamental Research Program, China (Grant Nos. JCYJ201500731154850923 and JCYJ20140417113430665), Shenzhen High-level Oversea Talent Program, China (Grant No. KQJSCX20160301144248), and the Nanshan Technology Research Fund, China (Grant No. KC2014JSQN0001A).

Corresponding Authors:  Yan-Chun Zhu     E-mail:  yc.zhu@siat.ac.cn

Cite this article: 

Shuo Li(李硕), Lei Wang(王磊), Yan-Chun Zhu(朱艳春), Jie Yang(杨洁), Yao-Qin Xie(谢耀钦), Nan Fu(付楠), Yi Wang(王乙), Song Gao(高嵩) Accuracy and effectiveness of self-gating signals in free-breathing three-dimensional cardiac cine magnetic resonance imaging 2016 Chin. Phys. B 25 128703

[1] Thattaliyath B D, Forsha D E, Stewart C, Barker P C A and Campbell M J 2015 Congenit Heart Dis. 10 371
[2] Ingle R R, Santos J M, Overall W R, McConnell M V, Hu B S and Nishimura D G 2015 Magnet. Reson. Med. 73 1764
[3] Elnakib A, Beache G M, Gimel'farb G and El-Baz A 2015 Int. J. Comput. Ass. Rad. 10 1299
[4] Saranathan M, Bayram E, Worters P W and Glockner J F 2012 Magn. Reson. Imaging 30 158
[5] Ng W H, Chan Y L, Sung J Y, Lee Y T, Lee S F and Chung S S C 2004 Magn. Reson. Mater. Phys. 16 211
[6] Peters D C, Ennis D B, Rohatgi P, Syed M A, McVeigh E R and Arai A E 2004 J. Magn. Reson. Imaging 20 411
[7] Makowski M R, Wiethoff A J, Jansen C H P, Uribe S, Parish V, Schuster A, Botnar R M, Bell A, Kiesewetter C, Razavi R, Schaeffter T and Greil G F 2012 J. Cardiovasc. Magn. R 14 1
[8] Jeong D, Schiebler M L, Lai P, Wang K, Vigen K K and Francois C J 2015 Int. J. Cardiovas. Imag. 31 851
[9] Larson A C, White R D, Laub G, McVeigh E R, Li D B and Simonetti O P 2004 Magnet. Reson. Med. 51 93
[10] Jahnke C, Paetsch I, Achenbach S, Schnackenburg B, Gebker R, Fleck E and Nagel E 2006 Radiology 239 71
[11] Palmucci S, Mauro L A, Coppolino M, Musumeci A G, Foti P V, Milone P and Ettorre G C 2010 Radiol. Med. 115 467
[12] Sievers B, Schrader S, Rehwald W, Hunold P, Barkhausen J and Erbel R 2011 Acta Cardiol. 66 349
[13] Sievers B, Schrader S, Hunold P, Barkhausen J and Erbel R 2011 Acta Cardiologica. 66 489
[14] Chen G J, Zhang Y J and Yang Y S 2013 Chin. Phys. B 22 124703
[15] Liu J, Nguyen T D, Zhu Y C, Spincemaille P, Prince M R, Weinsaft J W, Saloner D and Wang Y 2014 Plos One 9 e89315
[16] Liu J, Spincemaille P, Codella N C, Nguyen T D, Prince M R and Wang Y 2010 Magn. Reson. Med. 63 1230
[17] Zhu Y C, Liu J, Weinsaft J, Spincemaille P, Nguyen T D, Prince M R, Bao S L, Xie Y Q and Wang Y 2015 Biomed. Res. Int. 2015 819102
[18] Spincemaille P, Liu J, Nguyen T, Prince M R and Wang Y 2011 Magn. Reson. Imaging 29 861
[19] Tibiletti M, Paul J, Bianchi A, Wundrak S, Rottbauer W, Stiller D and Rasche V 2015 Magn. Reson. Med. 75 1324
[20] Wundrak S, Paul J, Ulrici J, Hell E, Geibel M A, Bernhardt P, Rottbauer W and Rasche V 2015 Magn. Reson. Med. 76 919
[21] Lehmkuhl L, Foldyna B, Von Aspern K, Lucke C, Grothoff M, Nitzsche S, Kempfert J, Haensig M, Rastan A, Walther T, Mohr F W and Gutberlet M 2013 Int. J. Cardiovas. Imag. 29 693
[22] Zhang J M, Guan Q, Tang L M, Liu T B, Liu H X, Huang X L and Si J F 2014 Chin. Phys B 23 018701
[23] Reis M S, Deus A P, Simoes R P, Aniceto I A V, Catai A M and Borghi-Silva A 2010 Rev. Bras. Fisioter. 14 106
[24] Ovchinnikiv K 2007 Psychophysiology 44 S49
[25] Elstad M, Walloe L, Holme N L A, Maes E and Thoresen M 2015 Eur. J. Appl. Physiol. 115 521
[26] Efimova E, Doering M, Richter S, Dinov B, Sommer P, Bollmann A, Fischer J, Garbade J, Hindricks G and Arya A 2015 Eur. Heart. J. 36 57
[27] Avitall B and Kalinski A 2015 Heart Rhythm. 12 2195
[28] Li W G, Zhang Z L, Li K A, Jin N, Zhang Y, Zhang T J, Miller F H and Larson A C 2015 Magnet. Reson. Med. 73 2249
[29] Jin N, Zhang Z L, Zhang L J, Lu G M and Larson A C 2011 Magnet. Reson. Med. 66 207
[30] Usman M, Atkinson D, Kolbitsch C, Schaeffter T and Prieto C 2015 J. Magn. Reson. Imaging 41 1521
[1] Response of HD-V2 radiochromic film to argon ions
Lei Cheng(程蕾), Zhe Zhang(张喆), Guiyun Liang(梁贵云), and Yutong Li(李玉同). Chin. Phys. B, 2021, 30(8): 080702.
[2] Parameter accuracy analysis of weak-value amplification process in the presence of noise
Jiangdong Qiu(邱疆冬), Zhaoxue Li(李兆雪), Linguo Xie(谢林果), Lan Luo(罗兰), Yu He(何宇), Changliang Ren(任昌亮), Zhiyou Zhang(张志友), and Jinglei Du(杜惊雷). Chin. Phys. B, 2021, 30(6): 064216.
[3] A transparent electromagnetic-shielding film based on one-dimensional metal-dielectric periodic structures
Ya-li Zhao(赵亚丽), Fu-hua Ma(马富花), Xu-feng Li(李旭峰), Jiang-jiang Ma(马江将), Kun Jia(贾琨), Xue-hong Wei(魏学红). Chin. Phys. B, 2018, 27(2): 027302.
[4] Performance study of aluminum shielded room for ultra-low-field magnetic resonance imaging based on SQUID: Simulations and experiments
Bo Li(李波), Hui Dong(董慧), Xiao-Lei Huang(黄小磊), Yang Qiu(邱阳), Quan Tao(陶泉), Jian-Ming Zhu(朱建明). Chin. Phys. B, 2018, 27(2): 020701.
[5] Accuracy design of ultra-low residual reflection coatingsfor laser optics
Huasong Liu(刘华松), Xiao Yang(杨霄), Lishuan Wang(王利栓), Hongfei Jiao(焦宏飞), Yiqin Ji(季一勤), Feng Zhang(张锋), D an Liu(刘丹丹), Chenghui Jiang(姜承慧), Yugang Jiang(姜玉刚), Deying Chen(陈德应). Chin. Phys. B, 2017, 26(7): 077801.
[6] Nonlinear profile order for three-dimensional hybrid radial acquisition applied to self-gated free-breathing cardiac cine MRI
Yanchun Zhu(朱艳春), Pascal Spincemaille, Jing Liu(刘静), Shuo Li(李硕), Thanh D Nguyen, Martin R Prince, Yaoqin Xie(谢耀钦), Yi Wang(王乙). Chin. Phys. B, 2017, 26(1): 018701.
[7] Calibration chain design based on integrating sphere transfer radiometer for SI-traceable on-orbit spectral radiometric calibration and its uncertainty analysis
Wei-Ning Zhao(赵维宁), Wei Fang(方伟), Li-Wei Sun(孙立微), Li-Hong Cui(崔立红), Yu-Peng Wang(王玉鹏). Chin. Phys. B, 2016, 25(9): 090701.
[8] Physical mechanism of mind changes and tradeoffs among speed, accuracy, and energy cost in brain decision making: Landscape, flux, and path perspectives
Han Yan(闫晗), Kun Zhang(张坤), Jin Wang(汪劲). Chin. Phys. B, 2016, 25(7): 078702.
[9] Improved kernel gradient free-smoothed particle hydrodynamics and its applications to heat transfer problems
Juan-Mian Lei(雷娟棉) and Xue-Ying Peng(彭雪莹). Chin. Phys. B, 2016, 25(2): 020202.
[10] Increasing the range accuracy of three-dimensional ghost imaging ladar using optimum slicing number method
Yang Xu (杨旭), Zhang Yong (张勇), Xu Lu (徐璐), Yang Cheng-Hua (杨成华), Wang Qiang (王强), Liu Yue-Hao (刘越豪), Zhao Yuan (赵远). Chin. Phys. B, 2015, 24(12): 124202.
[11] Novel wavelength-accurate InP-based arrayed waveguide grating
Pan Pan (潘盼), An Jun-Ming (安俊明), Wang Hong-Jie (王红杰), Wang Yue (王玥), Zhang Jia-Shun (张家顺), Wang Liang-Liang (王亮亮), Dai Hong-Qing (代红庆), Zhang Xiao-Guang (张晓光), Wu Yuan-Da (吴远大), Hu Xiong-Wei (胡雄伟). Chin. Phys. B, 2014, 23(4): 044210.
[12] Experimental study on supersonic film cooling on the surface of a blunt body in hypersonic flow
Fu Jia (付佳), Yi Shi-He (易仕和), Wang Xiao-Hu (王小虎), He Lin (何霖), Ge Yong (葛勇). Chin. Phys. B, 2014, 23(10): 104702.
[13] Theoretical description of improving measurement accuracy for incoherence Mie Doppler wind lidar
Du Jun (杜军), Ren De-Ming (任德明), Zhao Wei-Jiang (赵卫疆), Qu Yan-Chen (曲彦臣), Chen Zhen-Lei (陈振雷), Geng Li-Jie (耿利杰 ). Chin. Phys. B, 2013, 22(2): 024211.
[14] Effect of the nonlinearity of CCD in Fourier transform profilometry on spectrum overlapping and measurement accuracy
Qiao Nao-Sheng (乔闹生), Zou Bei-Ji (邹北骥). Chin. Phys. B, 2013, 22(1): 014203.
[15] Measurement accuracy analysis of the free carrier absorption determination of the electronic transport properties of silicon wafers
Zhang Xi-Ren(张希仁), Gao Chun-Ming(高椿明), Zhou Ying(周鹰), and Wang Zhan-Ping(王占平). Chin. Phys. B, 2011, 20(6): 068105.
No Suggested Reading articles found!