Single-shot grating-based x-ray differential phase contrast imaging with a modified analyzer grating

Chen-Xi Wei(卫晨希)1, Zhao Wu(吴朝)1, Faiz Wali1, Wen-Bin Wei(魏文彬)1, Yuan Bao(鲍园)2, Rong-Hui Luo(骆荣辉)1, Lei Wang(王磊)1, Gang Liu(刘刚)1, Yang-Chao Tian(田扬超)1
1. National Synchrotron Radiation Laboratory, University of Science and Technology of China, Hefei 230029, China;
2. Shanghai United Imaging Healthcare Co. Ltd., Shanghai 201807, China
Abstract  X-ray grating interferometer has attracted widely attention in the past years due to its capability in achieving x-ray phase contrast imaging with low brilliance source. However, the widely used phase stepping information extraction method reduces system stability and prolongs data acquisition time by several times compared with conventional x-ray absorption-based imaging. The mechanical stepping can be avoided by using a staggered grating, but at the cost of low vertical spatial resolution. In this paper, employing a modified staggered grating and the angular signal radiography, we proposed a single-shot grating-based x-ray differential phase contrast imaging with decent vertical spatial resolution. The theoretical framework was deduced and proved by numerical experiments. Absorption, phase, and scattering computed tomography can be performed without phase stepping. Therefore, we believe this fast and highly stable imaging method with decent resolution would be widely applied in x-ray grating-based phase contrast imaging.
Keywords:  x-ray phase contrast imaging      single-shot      decent resolution      staggered grating      computed tomography  
Received:  25 April 2017      Revised:  06 June 2017      Accepted manuscript online: 
PACS:  87.59.-e (X-ray imaging)  
  42.30.Rx (Phase retrieval)  
  87.57.Q- (Computed tomography)  
  41.50.+h (X-ray beams and x-ray optics)  
Fund: Project supported by the National Natural Science Foundation of China (Grant Nos. 11275204, 11475175, and 11405175), the China Postdoctoral Science Foundation (Grant No. 2017M612097), and the Fundamental Research Funds for the Central Universities (Grant No. WK2310000065).
Corresponding Authors:  Zhao Wu, Yang-Chao Tian     E-mail:  wuzhao@ustc.edu.cn;ychtian@ustc.edu.cn

Cite this article: 

Chen-Xi Wei(卫晨希), Zhao Wu(吴朝), Faiz Wali, Wen-Bin Wei(魏文彬), Yuan Bao(鲍园), Rong-Hui Luo(骆荣辉), Lei Wang(王磊), Gang Liu(刘刚), Yang-Chao Tian(田扬超) Single-shot grating-based x-ray differential phase contrast imaging with a modified analyzer grating 2017 Chin. Phys. B 26 108701

[1] Pfeiffer F, Herzen J, Willner M, Chabior M, Auweter S, Reiser M and Bamberg F 2013 Z. Für Med. Phys. 23 176
[2] Momose A, Yashiro W, Kido K, Kiyohara J, Makifuchi C, Ito T, Nagatsuka S, Honda C, Noda D, Hattori T, Endo T, Nagashima M and Tanaka J 2014 Philos. Trans. R. Soc. Math. Phys. Eng. Sci. 372 20130023
[3] Pfeiffer F, Kottler C, Bunk O and David C 2007 Phys. Rev. Lett. 98 108105
[4] Chen G H, Bevins N, Zambelli J and Qi Z 2010 Opt. Express 18 12960
[5] Huang Z F, Kang K J, Zhang L, Chen Z Q, Ding F, Wang Z T and Fang Q G 2009 Phys. Rev. A 79 013815
[6] Momose A, Kawamoto S, Koyama I and Suzuki Y 2004 AIP Conf. Proc. 716 pp. 156-159
[7] Kagias M, Wang Z, Villanueva-Perez P, Jefimovs K and Stampanoni M 2016 Phys. Rev. Lett. 116 093902
[8] Zanette I, Bech M, Pfeiffer F and Weitkamp T 2011 Appl. Phys. Lett. 98 094101
[9] Bevins N, Zambelli J, Li K, Qi Z and Chen G H 2011 Med. Phys. 39 424
[10] Huang Z F, Chen Z Q, Zhang L, Kang K J, Ding F, Wang Z T and Ma H Z 2010 Opt. Express 18 10222
[11] Wu Z, Gao K, Chen J, Wang D, Wang S, Chen H, Bao Y, Shao Q, Wang Z, Zhang K, Zhu P and Wu Z 2015 Med. Phys. 42 741
[12] Wu Z, Gao K, Wang Z, Ge X, Chen J, Wang D and Pan Z 2013 Med. Phys. 40 031911
[13] Momose A P 2003 Opt. Express 11 2303
[14] Weitkamp T, Diaz A and David C 2005 Opt. Express 13 6296
[15] Miao H, Chen L, Bennett E E, Adamo N M, Gomella A A, Deluca A M, Patel A, Morgan N Y and Wen H 2013 Proc. Natl. Acad. Sci. 110 19268
[16] Zhu P, Zhang K, Wang Z, Liu Y, Liu X, Wu Z, A. McDonald S, Marone F and Stampanoni M 2010 Proc. Natl. Acad. Sci. 107 13576
[17] Nesterets Y I and Wilkins S W 2008 Opt. Express 16 5849
[18] Ge Y, Li K, Garrett J and Chen G H 2014 Opt. Express 22 14246
[19] Wang Z, Gao K, Wang D, Wu Z, Chen H, Wang S and Wu Z 2014 Opt. Lett. 39 877
[20] Li P, Zhang K, Bao Y, Ren Y, Ju Z, Wang Y, He Q, Zhu Z, Huang W, Yuan Q and Zhu P 2016 Opt. Express 24 5829
[21] Wang Z T, Kang K J, Huang Z F and Chen Z Q 2009 Appl. Phys. Lett. 95 094105
[22] Bao Y, Shao Q, Hu R, Wang S, Gao K, Wang Y, Tian Y and Zhu P 2016 Radiat. Phys. Chem. 137 260
[23] Pfeiffer F, Bech M, Bunk O, Donath T, Henrich B, Kraft P and David C 2009 J. Appl. Phys. 105 102006
[1] X-ray phase-sensitive microscope imaging with a grating interferometer: Theory and simulation
Jiecheng Yang(杨杰成), Peiping Zhu(朱佩平), Dong Liang(梁栋), Hairong Zheng(郑海荣), and Yongshuai Ge(葛永帅). Chin. Phys. B, 2022, 31(9): 098702.
[2] High-resolution bone microstructure imaging based on ultrasonic frequency-domain full-waveform inversion
Yifang Li(李义方), Qinzhen Shi(石勤振), Ying Li(李颖), Xiaojun Song(宋小军), Chengcheng Liu(刘成成), Dean Ta(他得安), and Weiqi Wang(王威琪). Chin. Phys. B, 2021, 30(1): 014302.
[3] Single-shot measurement of THz pulses
Lei Yang(杨磊), Lei Hou(侯磊), Chengang Dong(董陈岗), Wei Shi(施卫). Chin. Phys. B, 2020, 29(5): 057803.
[4] Shifting curves based on the detector integration effect for x-ray phase contrast imaging
Jun Yang(杨君), Jin-Chuan Guo(郭金川), Yao-Hu Lei(雷耀虎), Ming-Hao Yi(易明皓), Li Chen(陈力). Chin. Phys. B, 2017, 26(2): 028701.
[5] Optimization-based image reconstruction in x-ray computed tomography by sparsity exploitation of local continuity and nonlocal spatial self-similarity
Han-Ming Zhang(张瀚铭), Lin-Yuan Wang(王林元), Lei Li(李磊), Bin Yan(闫镔), Ai-Long Cai(蔡爱龙), Guo-En Hu(胡国恩). Chin. Phys. B, 2016, 25(7): 078701.
[6] Cosine fitting radiography and computed tomography
Li Pan-Yun (李盼云), Zhang Kai (张凯), Huang Wan-Xia (黄万霞), Yuan Qing-Xi (袁清习), Wang Yan (王研), Ju Zai-Qiang (鞠在强), Wu Zi-Yu (吴自玉), Zhu Pei-Ping (朱佩平). Chin. Phys. B, 2015, 24(6): 068704.
[7] Fast parallel algorithm for three-dimensional distance-driven model in iterative computed tomography reconstruction
Chen Jian-Lin (陈建林), Li Lei (李磊), Wang Lin-Yuan (王林元), Cai Ai-Long (蔡爱龙), Xi Xiao-Qi (席晓琦), Zhang Han-Ming (张瀚铭), Li Jian-Xin (李建新), Yan Bin (闫镔). Chin. Phys. B, 2015, 24(2): 028703.
[8] Enhancement of photoacoustic tomography in the tissue with speed-of-sound variance using ultrasound computed tomography
Cheng Ren-Xiang (程任翔), Tao Chao (陶超), Liu Xiao-Jun (刘晓峻). Chin. Phys. B, 2015, 24(11): 114301.
[9] Investigation of noise properties in grating-based x-ray phase tomography with reverse projection method
Bao Yuan (鲍园), Wang Yan (王研), Gao Kun (高昆), Wang Zhi-Li (王志立), Zhu Pei-Ping (朱佩平), Wu Zi-Yu (吴自玉). Chin. Phys. B, 2015, 24(10): 108702.
[10] Scatter correction method for cone-beam CT based on interlacing-slit scan
Huang Kui-Dong (黄魁东), Zhang Hua (张华), Shi Yi-Kai (史仪凯), Zhang Liang (张亮), Xu Zhe (徐哲). Chin. Phys. B, 2014, 23(9): 098106.
[11] Formation of multifunctional Fe3O4/Au composite nanoparticles for dual-mode MR/CT imaging applications
Hu Yong (胡勇), Li Jing-Chao (李静超), Shen Ming-Wu (沈明武), Shi Xiang-Yang (史向阳). Chin. Phys. B, 2014, 23(7): 078704.
[12] Detection of the spatiotemporal field of a single-shot terahertz pulse based on spectral holography
Wang Xiao-Lei (王晓雷), Fei Yang (费扬), Li Lu-Jie (李璐杰), Wang Qiang (王强), Zhu Zhu-Qing (朱竹青). Chin. Phys. B, 2014, 23(6): 064202.
[13] New terahertz dispersive device for single-shot spectral measurements of terahertz pulse
Wu Yin-Zhong (巫殷忠), Zhao Zong-Qing (赵宗清), Gu Yu-Qiu (谷渝秋), Wei Lai (魏来), Cao Lei-Feng (曹磊峰). Chin. Phys. B, 2014, 23(12): 125202.
[14] Penetrating view of nano-structures in Aleochara verna spermatheca and flagellum by hard X-ray microscopy
Zhang Kai (张凯), Li Dee (李德娥), Hong You-Li (洪友丽), Zhu Pei-Ping (朱佩平), Yuan Qing-Xi (袁清习), Huang Wan-Xia (黄万霞), Gao Kun (高昆), Zhou Hong-Zhang (周红章), Wu Zi-Yu (吴自玉). Chin. Phys. B, 2013, 22(7): 076801.
[15] Multiple helical scans and the reconstruction of over FOV-sized objects in cone-beam CT
Han Yu(韩玉), Yan Bin(闫镔), Li Lei(李磊), Yu Chao-Qun(宇超群), Li Jian-Xin(李建新), and Bao Shang-Lian(包尚联) . Chin. Phys. B, 2012, 21(6): 068701.
No Suggested Reading articles found!