Scatter correction method for cone-beam CT based on interlacing-slit scan |
Huang Kui-Dong (黄魁东)a, Zhang Hua (张华)b, Shi Yi-Kai (史仪凯)b, Zhang Liang (张亮)a, Xu Zhe (徐哲)a |
a Key Laboratory of Contemporary Design and Integrated Manufacturing Technology of the Ministry of Education, Northwestern Polytechnical University, Xi'an 710072, China; b School of Mechanical Engineering, Northwestern Polytechnical University, Xi'an 710072, China |
|
|
Abstract Cone-beam computed tomography (CBCT) has the notable features of high efficiency and high precision, and is widely used in areas such as medical imaging and industrial non-destructive testing. However, the presence of the ray scatter reduces the quality of CT images. By referencing the slit collimation approach, a scatter correction method for CBCT based on the interlacing-slit scan is proposed. Firstly, according to the characteristics of CBCT imaging, a scatter suppression plate with interlacing slits is designed and fabricated. Then the imaging of the scatter suppression plate is analyzed, and a scatter correction calculation method for CBCT based on the image fusion is proposed, which can splice out a complete set of scatter suppression projection images according to the interlacing-slit projection images of the left and the right imaging regions in the scatter suppression plate, and simultaneously complete the scatter correction within the flat panel detector (FPD). Finally, the overall process of scatter suppression and correction is provided. The experimental results show that this method can significantly improve the clarity of the slice images and achieve a good scatter correction.
|
Received: 04 November 2013
Revised: 03 March 2014
Accepted manuscript online:
|
PACS:
|
81.70.Tx
|
(Computed tomography)
|
|
87.57.Q-
|
(Computed tomography)
|
|
87.64.Bx
|
(Electron, neutron and x-ray diffraction and scattering)
|
|
87.57.C-
|
(Image quality)
|
|
Fund: Project supported by the National Science and Technology Major Project of the Ministry of Industry and Information Technology of China (Grant No. 2012ZX04007021), the Young Scientists Fund of the National Natural Science Foundation of China (Grant No. 51105315), the Natural Science Basic Research Program of Shaanxi Province, China (Grant No. 2013JM7003), and the Northwestern Polytechnical University Foundation for Fundamental Research, China (Grant Nos. 3102014KYJD022 and JC20120226). |
Corresponding Authors:
Huang Kui-Dong
E-mail: kdhuang@nwpu.edu.cn
|
Cite this article:
Huang Kui-Dong (黄魁东), Zhang Hua (张华), Shi Yi-Kai (史仪凯), Zhang Liang (张亮), Xu Zhe (徐哲) Scatter correction method for cone-beam CT based on interlacing-slit scan 2014 Chin. Phys. B 23 098106
|
[1] |
Luo Z Y, Yang X Q, Meng Y Z and Deng Y 2010 Acta Phys. Sin. 59 8237 (in Chinese)
|
[2] |
Wang X C, Yan B, Liu H K, Li L, Wei X and Hu G E 2013 Acta Phys. Sin. 62 098702 (in Chinese)
|
[3] |
Guo X L, Tang Z W, Zhang H and Hu G S 2010 J. Tsinghua Univ. (Sci. & Tech.) 50 450
|
[4] |
Ding G X, Duggan D M and Coffey C W 2007 Phys. Med. Biol. 52 1595
|
[5] |
Morin O, Gillis A, Chen J, Aubin M, Bucci M K, Roach M 3rd and Pouliot J 2006 Med. Dosim. 31 51
|
[6] |
Kwan A L, Boone J M and Shah N 2005 Med. Phys. 32 2967
|
[7] |
Siewerdsen J H and Jaffrey D A 2001 Med. Phys. 28 220
|
[8] |
Boone J M 1986 Med. Phys. 13 319
|
[9] |
Neitzel U 1992 Med. Phys. 19 475
|
[10] |
Graham S A, Moseley D J, Siewerdsen J H and Jaffray D A 2007 Med. Phys. 34 2691
|
[11] |
Coates A M, Dixon A M and Huckle J 2012 Radiography 18 250
|
[12] |
Li Q L, Yan B, Li L, Sun H S and Zhang F 2012 High Power Laser and Particle Beams 24 2235
|
[13] |
Naimuddin S, Hasegawa B H and Mistretta C A 1987 Med. Phys. 14 330
|
[14] |
Deng Z S, Ni D P, Jiang D Z, Zhao Z H, Xiao F and Xu L J 1997 J. Biomed. Eng. 14 137 (in Chinese)
|
[15] |
Jia P X, Zhang F, Yan B and Bao S L 2010 Chin. Phys. B 19 087802
|
[16] |
Mettivier G, Lanconelli N, Meo S L and Russo P 2012 IEEE Trans. Nucl. Sci. 59 2008
|
[17] |
Baer M and Kachelrie M 2012 Phys. Med. Biol. 57 6849
|
[18] |
Ning R, Tang X Y and Conover D 2004 Med. Phys. 31 1195
|
[19] |
Schörner K, Goldammer M and Stephan J 2011 Nucl. Instr. Meth. Phys. Res. B 269 292
|
[20] |
Zhang D H, Hu D C, Huang K D and Kong Y M 2009 Chin. Mech. Eng. 20 639 (in Chinese)
|
[21] |
Niu T Y and Zhu L 2011 Med. Phys. 38 6027
|
[22] |
Dong X, Jia X, Niu T Y and Zhu L 2012 Proc. SPIE Medical Imaging 2012: Physics of Medical Imaging, February 4-9, 2012 San Diego, USA, p. 831319
|
[23] |
Wang J, Mao W and Solberg T 2010 Med. Phys. 37 5792
|
[24] |
Ouyang L, Song K and Wang J 2013 Med. Phys. 40 071903
|
[25] |
Xie S P and Luo L M 2011 Acta Electron. Sin. 39 1708 (in Chinese)
|
[26] |
Zhu L, Bennett N R and Fahrig R 2006 IEEE Trans. Med. Imaging 25 1573
|
[27] |
Gao H W, Fahrig R, Bennett N R, Sun M S, Star-Lack J and Zhu L 2010 Med. Phys. 37 934
|
[28] |
Wils P, Létang J M and Bruandet J P 2012 J. Electron. Imaging 21 021113
|
[29] |
Zhang H, Huang K D, Shi Y K and Li M J 2012 CT Theory and Applications 21 247 (in Chinese)
|
[30] |
Huang K D, Zhang D H, Li M J and Zhang H 2013 Acta Phys. Sin. 62 210702 (in Chinese)
|
No Suggested Reading articles found! |
|
|
Viewed |
|
|
|
Full text
|
|
|
|
|
Abstract
|
|
|
|
|
Cited |
|
|
|
|
Altmetric
|
blogs
Facebook pages
Wikipedia page
Google+ users
|
Online attention
Altmetric calculates a score based on the online attention an article receives. Each coloured thread in the circle represents a different type of online attention. The number in the centre is the Altmetric score. Social media and mainstream news media are the main sources that calculate the score. Reference managers such as Mendeley are also tracked but do not contribute to the score. Older articles often score higher because they have had more time to get noticed. To account for this, Altmetric has included the context data for other articles of a similar age.
View more on Altmetrics
|
|
|