Please wait a minute...
Chin. Phys. B, 2011, Vol. 20(3): 030702    DOI: 10.1088/1674-1056/20/3/030702
GENERAL Prev   Next  

Simulation study of a magnetocardiogram based on a virtual heart model: effect of a cardiac equivalent source and a volume conductor

Shou Guo-Fa(寿国法)a), Xia Ling(夏灵) a)†, Ma Ping(马平)b), Tang Fa-Kuan(唐发宽)c)‡, and Dai Ling(戴灵)a)
a Department of Biomedical Engineering, Key Laboratory for Biomedical Engineering of the Ministry of Education, Zhejiang University, Hangzhou 310027, China; b Department of Physics, State Key Laboratory for Artificial Microstructure and Mesoscopic Physics, Peking University, Beijing 100871, China; c Department of Cardiology, the 309th Hospital of the People's Liberation Army, Beijing 100091, China
Abstract  In this paper, we present a magnetocardiogram (MCG) simulation study using the boundary element method (BEM) and based on the virtual heart model and the realistic human volume conductor model. The different contributions of cardiac equivalent source models and volume conductor models to the MCG are deeply and comprehensively investigated. The single dipole source model, the multiple dipoles source model and the equivalent double layer (EDL) source model are analysed and compared with the cardiac equivalent source models. Meanwhile, the effect of the volume conductor model on the MCG combined with these cardiac equivalent sources is investigated. The simulation results demonstrate that the cardiac electrophysiological information will be partly missed when only the single dipole source is taken, while the EDL source is a good option for MCG simulation and the effect of the volume conductor is smallest for the EDL source. Therefore, the EDL source is suitable for the study of MCG forward and inverse problems, and more attention should be paid to it in future MCG studies.
Keywords:  virtual heart model      magnetocardiogram      dipole source      equivalent double layer source      volume conductor model  
Received:  05 August 2010      Revised:  08 September 2010      Accepted manuscript online: 
PACS:  07.05.Tp (Computer modeling and simulation)  
Fund: Project supported by the State Key Development Program for Basic Research of China (Grant Nos. 2007CB512100 and 2006CB601007), the National Natural Science Foundation of China (Grant No. 10674006), the National High Technology Research and Development Program of China (Grant No. 2007AA03Z238), China Postdoctoral Science Foundation (Grant No. 20090461376), and the Fundamental Research Funds for the Central Universities (Grant No. KYJD09001).

Cite this article: 

Shou Guo-Fa(寿国法), Xia Ling(夏灵), Ma Ping(马平), Tang Fa-Kuan(唐发宽), and Dai Ling(戴灵) Simulation study of a magnetocardiogram based on a virtual heart model: effect of a cardiac equivalent source and a volume conductor 2011 Chin. Phys. B 20 030702

[1] Dos Santos R W, Kosch O, Steinhoff U, Bauer S, Trahms L and Koch H 2004 J. Electrocardiol. 37 123
[2] Lim H K, Chung N, Ko Y G, Lee Y H and Park Y K 2009 IEEE Trans. Magn. 45 2890
[3] Im B U, Kwon S S, Kim K, Lee Y H, Park Y K, Youn C H and Shim E B 2008 Prog. Biophys. Mol. Biol. 96 339
[4] Smith F E,Langley P, van Leeuwen P, Hailer B, Trahms L, Steinhoff U, Bourke J P and Murray A 2007 Anadolu Kardiyoloji. Derg. 7 20
[5] Yamada S and Yamaguchi I 2005 Intern. Med. 44 1
[6] Fenici R, Brisinda D and Meloni A M 2005 Expert Rev. Mol. Diagn. 5 291
[7] Jurkko R, Mantynen V, Tapanainen J M, Montonen J, Vaananen H, Parikka H and Toivonen L 2009 Europace 11 169
[8] Wang C N, Li W Q, Zhang J and Fang P H 2006 Chin. J. Med. 41 9 (in Chinese)
[9] Koch H 2001 IEEE Trans. Appl. Supercond. 11 49
[10] Liu X Y, Xie F X, Meng S C, Ma P, Yang T, Nie R J, Wang S Z, Wang F R and Dai Y D 2003 Acta Phys. Sin. 52 2580 (in Chinese)
[11] Zhao L, Chen G H, Zhang L H, Huang X G, Zhai G J, Li J W, Tang Y L and Feng J 2004 Acta Phys. Sin. 53 4420 (in Chinese)
[12] Liu X Y, Xie B Q, Dai Y D, Wang F R, Li Z Z, Ma P, Xie F X, Yang T and Nie R J 2005 Acta Phys. Sin. 54 1937 (in Chinese)
[13] Zhang S L, Wang Y L, Wang H W, Jiang S Q and Xie X M 2009 Phy. Med. Biol. 54 4793
[14] Ogata K, Kandori A, Miyashita T, Tsukada K, Nakatani S, Shimizu W, Kanzaki H, Miyatake K, Yamada S, Watanabe S and Yamaguchi I 2006 Int. J. Cardiovasc. Imag. 22 581
[15] Ramon C, Czapski P, Haueisen J, Huntsman L L, Nowak H, Bardy G H, Leder U, Kim Y and Nelson J A 1998 IEEE Trans. Biomed. Eng. 45 1323
[16] Geselowitz D B 1980 IEEE Trans. Magn. 16 812
[17] Czapski P, Ramon C, Huntsman L L, Bardy G H and Kim Y 1996 Phys. Med. Biol. 41 1247
[18] Wang Q, Hua N, Lu H, Ma P, Tang F K and Tang X Z 2010 Chin. Phys. B 19 080601
[19] Haueisen J, Schreiber J, Brauer H and Knosche T R 2002 IEEE Trans. Magn. 38 1045
[20] Wang Q, Ma P, Hua N, Lu H, Tang X Z and Tang F K 2010 Acta Phys. Sin. 59 2882 (in Chinese)
[21] De Melis M, Tanaka K and Uchikawa Y 2010 IEEE Trans. Magn. 46 1203
[22] Wang Q, Ma P, Lu H, Tang X Z, Hua N and Tang F K 2009 Chin. Phys. B 18 5566
[23] van Oosterom A 2003 Int. J. Bioelectromagnetism 5 211
[24] Zhang L, Jiang S Q, Wang L, Luo M and Wang L M 2005 Modern Scientific Instruments 4 31 (in Chinese)
[25] van Oosterom A, Oostendorp T F, Huiskamp G J and Terbrake H J M 1990 Circ. Res. 67 1503
[26] Shou G F, Xia L, Jiang M F, Wei Q, Liu F and Crozier S 2009 IEEE Trans. Biomed. Eng. 56 1454
[27] Gulrajani R M 1988 Crc. Critical. Rev. Biomed. Eng. 16 1
[28] Yamashita Y and Geselowitz D B 1985 IEEE Trans. Biomed. Eng. 32 964
[29] Xia L, Huo M, Wei Q, Liu F and Crozier S 2006 Methods Inf. Med. 45 564
[30] Lu W X and Xia L 1996 IEEE Trans. Biomed. Eng. 43 211 endfootnotesize
[1] Using Helmholtz resonator arrays to improve dipole transmission efficiency in waveguide
Liwei Wang(王力维), Li Quan(全力), Feng Qian(钱枫), Xiaozhou Liu(刘晓宙). Chin. Phys. B, 2019, 28(9): 094301.