A normalized data-reusing least-mean-square algorithm of noise cancellation for magnetocardiography

孔祥燕, 王会武, 田野, 黄旭光, 张利华, 任育峰, 陈赓华, 杨乾声   

  1. Institute of Physics, Chinese Academy of Sciences, Beijing 100080, China
  • 收稿日期:2004-04-16 修回日期:2004-06-08 出版日期:2005-06-20 发布日期:2005-06-20
  • 基金资助:
    Project supported by Chinese Academy of Sciences (Grant No KJCX2-W4) and the Ministry of Science and Technology of China (Grant No 2002AA306412 and NKBRSF-G1999064609).

A normalized data-reusing least-mean-square algorithm of noise cancellation for magnetocardiography

Kong Xiang-Yan (孔祥燕), Wang Hui-Wu (王会武), Tian Ye (田野), Huang Xu-Guang (黄旭光), Zhang Li-Hua (张利华), Ren Yu-Feng (任育峰), Chen Geng-Hua (陈赓华), Yang Qian-Sheng (杨乾声)   

  1. Institute of Physics, Chinese Academy of Sciences, Beijing 100080, China
  • Received:2004-04-16 Revised:2004-06-08 Online:2005-06-20 Published:2005-06-20
  • Supported by:
    Project supported by Chinese Academy of Sciences (Grant No KJCX2-W4) and the Ministry of Science and Technology of China (Grant No 2002AA306412 and NKBRSF-G1999064609).

摘要: An adaptive filter based on Data-Reusing Least-Mean-Square algorithm has been proposed and applied in magnetocardiography (MCG) to suppress ambient noise. Numerical simulation studies indicate that the adaptive filter is a powerful noise suppresser for correlated interferences, especially for those with amplitude changing and time delay. Also the filter has a wide frequency bandwidth. With this filter, the signal-to-noise ratio of an MCG is improved to the intrinsic noise level. The periodic average method for further improvement of the noise level in MCG is also discussed.

Abstract: An adaptive filter based on Data-Reusing Least-Mean-Square algorithm has been proposed and applied in magnetocardiography (MCG) to suppress ambient noise. Numerical simulation studies indicate that the adaptive filter is a powerful noise suppresser for correlated interferences, especially for those with amplitude changing and time delay. Also the filter has a wide frequency bandwidth. With this filter, the signal-to-noise ratio of an MCG is improved to the intrinsic noise level. The periodic average method for further improvement of the noise level in MCG is also discussed.

Key words: magnetocardiography, noise suppression

中图分类号:  (Cardiac dynamics)

  • 87.19.Hh
84.30.Vn (Filters) 84.40.Ua (Telecommunications: signal transmission and processing; communication satellites) 02.60.-x (Numerical approximation and analysis) 02.70.Rr (General statistical methods)