Imaging plate scanners calibration and the attenuation behavior of imaging plate signals

Nan Bo(薄楠)1,2 and Nai-Yan Wang(王乃彦)2,†   

  1. 1 College of Nuclear Science and Technology, Beijing Normal University, Beijing 100875, China;
    2 China Institute of Atomic Energy, Beijing 102413, China
  • 收稿日期:2024-04-24 修回日期:2024-04-26 接受日期:2024-04-12 出版日期:2024-06-18 发布日期:2024-06-18
  • 通讯作者: Nai-Yan Wang E-mail:wangny@bnu.edu.cn
  • 基金资助:
    This work was supported by the Nuclear Industry Academician Fund, the National Natural Science Foundation of China (Grant No. 1220051312), and Young Talents Fund of China National Nuclear Corporation (Grant No. FY212406000901).

Imaging plate scanners calibration and the attenuation behavior of imaging plate signals

Nan Bo(薄楠)1,2 and Nai-Yan Wang(王乃彦)2,†   

  1. 1 College of Nuclear Science and Technology, Beijing Normal University, Beijing 100875, China;
    2 China Institute of Atomic Energy, Beijing 102413, China
  • Received:2024-04-24 Revised:2024-04-26 Accepted:2024-04-12 Online:2024-06-18 Published:2024-06-18
  • Contact: Nai-Yan Wang E-mail:wangny@bnu.edu.cn
  • Supported by:
    This work was supported by the Nuclear Industry Academician Fund, the National Natural Science Foundation of China (Grant No. 1220051312), and Young Talents Fund of China National Nuclear Corporation (Grant No. FY212406000901).

摘要: Based on previously reported work, we propose a new method for calibrating image plate (IP) scanners, offering greater flexibility and convenience, which can be extended to the calibration tasks of various scanner models. This method was applied to calibrate the sensitivity of a GE Typhoon FLA 7000 scanner. Additionally, we performed a calibration of the spontaneous signal attenuation behavior for BAS-MS, BAS-SR, and BAS-TR type IPs under the 20$\pm$1$^\circ$C environmental conditions, and observed significant signal carrier diffusion behavior in BAS-MS IP. The calibration results lay a foundation for further research on the interaction between ultra-short, ultra-intense lasers and matter.

关键词: image plate, scanner, ultra-short ultra-intense lasers

Abstract: Based on previously reported work, we propose a new method for calibrating image plate (IP) scanners, offering greater flexibility and convenience, which can be extended to the calibration tasks of various scanner models. This method was applied to calibrate the sensitivity of a GE Typhoon FLA 7000 scanner. Additionally, we performed a calibration of the spontaneous signal attenuation behavior for BAS-MS, BAS-SR, and BAS-TR type IPs under the 20$\pm$1$^\circ$C environmental conditions, and observed significant signal carrier diffusion behavior in BAS-MS IP. The calibration results lay a foundation for further research on the interaction between ultra-short, ultra-intense lasers and matter.