中国物理B ›› 2023, Vol. 32 ›› Issue (10): 100703-100703.doi: 10.1088/1674-1056/acca08
Jian Liu(刘建)1, Dongming Wang(王东明)1, Yuecheng Fu(甫跃成)1, Zhongbao Li(李忠宝)1, Han Yi(易晗)2, and Longtao Yi(易龙涛)1,†
Jian Liu(刘建)1, Dongming Wang(王东明)1, Yuecheng Fu(甫跃成)1, Zhongbao Li(李忠宝)1, Han Yi(易晗)2, and Longtao Yi(易龙涛)1,†
摘要: A scintillator detector consisting of a LaBr3(Ce) (0.5%) scintillator, a photomultiplier tube (PMT), and an oscilloscope were used to study the neutron sensitivities of the LaBr3(Ce) scintillator at the China Spallation Neutron Source (CSNS) Back-n white neutron source in the double-bunch and single-bunch operation modes, respectively. Under the two operational modes, the relative neutron sensitivity curves of the LaBr3(Ce) scintillator in the energy regions of 1-20 MeV and 0.5-20 MeV were obtained for the first time. In the energy range of 1-20 MeV, the two curves were nearly identical. However the relative neutron sensitivity uncertainties of the double-bunch experiment were higher than those of the single-bunch experiment. The above results indicated that the single-bunch experiment's neutron sensitivity curve has a lower minimum measurable energy than the double-bunch experiment. Above the minimum measurable energy of the double-bunch experiment, there is little difference between the measured relative neutron sensitivity curves of the single-bunch and double-bunch experiments of the LaBr3(Ce) scintillator and those of other scintillators with a similar neutron response signal intensity.
中图分类号: (X- and γ-ray sources, mirrors, gratings, and detectors)