Xiao-Yu Xing, Jing-Qiang Han, Wen-Tao Lu, Zhi-Qiang Zhang, Yang Wang, Shu Liu, Na Li, Yin-Jun Wang, Zi-Yan Wu, Xin-Yu Liu, Sheng-Kai Xia, Ke-Qi Shi, Bo Yang, Chen-Lu Bu, Guan-Xiang Du. Integrated flexible microwave device for magnetic material scanning using nitrogen-vacancy centerJ. Chin. Phys. B, 2026, 35(9): 094215.
| Xiao-Yu Xing, Jing-Qiang Han, Wen-Tao Lu, Zhi-Qiang Zhang, Yang Wang, Shu Liu, Na Li, Yin-Jun Wang, Zi-Yan Wu, Xin-Yu Liu, Sheng-Kai Xia, Ke-Qi Shi, Bo Yang, Chen-Lu Bu, Guan-Xiang Du. Integrated flexible microwave device for magnetic material scanning using nitrogen-vacancy centerJ. Chin. Phys. B, 2026, 35(9): 094215. |
Integrated flexible microwave device for magnetic material scanning using nitrogen-vacancy center
-
Abstract
We design a flexible microwave transmission line based on a splitting-combining structure and integrate it with a fiber-diamond probe to form a unified quantum probe, providing a resonant driving microwave magnetic field. This device utilizes a polyimide (PI) flexible substrate, which is conformable. A through-hole is designed at the center of the ring to accommodate fiber insertion, enabling the sensing unit to approach the sample under test. Experimental results demonstrate that the device achieves a 394~\rm MHz bandwidth, with a microwave efficiency of 4.64~\rm G/\sqrt W , balancing broadband with a strong near-field microwave magnetic field. Finally, we characterize the spatial distributions of the static magnetic-field magnitude and the angle between the static magnetic field and the selected NV axis on the surface of the kagome magnet TmMn_6Sn_6, demonstrating the application potential of the proposed probe for magnetic-material characterization. -