Print ISSN:1674-1056  |  Online ISSN:2058-3834  |  CN:11-5639/O4
  • Cite this article:

    Jiajie Li, Xiujie Fang, Renjie Li, Baodong Chen, Yueyang Zhai, Ying Liu. A miniaturized spin-exchange relaxation-free atomic magnetometer based on uniform light fieldJ. Chin. Phys. B, 2023, 32(5): 053201.
    Jiajie Li, Xiujie Fang, Renjie Li, Baodong Chen, Yueyang Zhai, Ying Liu. A miniaturized spin-exchange relaxation-free atomic magnetometer based on uniform light fieldJ. Chin. Phys. B, 2023, 32(5): 053201.
  • A miniaturized spin-exchange relaxation-free atomic magnetometer based on uniform light field

    • We experimentally study the dynamic characteristics of a miniaturized spin-exchange relaxation-free (SERF) magnetometer based on uniform light field. The ceramic ferrule is used to expand the Gaussian beam to improve light intensity uniformity, while the volume of the sensor is also reduced. This scheme makes the magnetometer have better sensitivity when the detected light intensity is less than 3.16 mW/cm 2 at 120 °C. When the temperature rises to 150 °C the sensitivity under the action of uniform light field is 18.5 fT/Hz 1/2. The bandwidth of the sensor remains at the original level and meets application needs. The proposed structure improves transverse polarization uniformity within the miniaturized sensor, which is ideal for the magnetoencephalography and magnetocardiography imaging systems.
    • Article Text

    • loading

    Catalog

      /

      DownLoad:  Full-Size Img  PowerPoint
      Return
      Return