Print ISSN:1674-1056  |  Online ISSN:2058-3834  |  CN:11-5639/O4
  • Cite this article:

    Hong-Tai Xie, Bin Chen, Jin-Bao Long, Chun Xue, Luo-Kan Chen, Shuai Chen. Calibration of a compact absolute atomic gravimeterJ. Chin. Phys. B, 2020, 29(9): 093701.
    Hong-Tai Xie, Bin Chen, Jin-Bao Long, Chun Xue, Luo-Kan Chen, Shuai Chen. Calibration of a compact absolute atomic gravimeterJ. Chin. Phys. B, 2020, 29(9): 093701.
  • Calibration of a compact absolute atomic gravimeter

    • Compact atomic gravimeters are the potential next generation precision instruments for gravity survey from fundamental research to broad field applications. We report the calibration results of our home build compact absolute atomic gravimeter USTC-AG02 at Changping Campus, the National Institute of Metrology (NIM), China in January 2019. The sensitivity of the atomic gravimeter reaches 35.5 μGal/√Hz (1 μGal=1×10-8 m/s2) and its long-term stability reaches 0.8 μGal for averaging over 4000 seconds. Considering the statistical uncertainty, the dominant instrumental systematic errors and environmental effects are evaluated and corrected within a total uncertainty (2σ) of 15.3 μGal. After compared with the reference g value given by the corner cube gravimeter NIM-3A, the atomic gravimeter USTC-AG02 reaches the degree of equivalence of 3.7 μGal.
    • Article Text

    • loading

    Catalog

      /

      DownLoad:  Full-Size Img  PowerPoint
      Return
      Return