Cite this article:
Min Wei, Qing-Hao Ye, Rui-Feng Kan, Bing Chen, Chen-Guang Yang, Zhen-Yu Xu, Xiang Chen, Jun Ruan, Xue-Li Fan, Wei Wang, Mai Hu, Jian-Guo Liu. Spectroscopy system based on a single quantum cascade laser for simultaneous detection of CO and CO2J. Chin. Phys. B, 2016, 25(9): 094210.
| Min Wei, Qing-Hao Ye, Rui-Feng Kan, Bing Chen, Chen-Guang Yang, Zhen-Yu Xu, Xiang Chen, Jun Ruan, Xue-Li Fan, Wei Wang, Mai Hu, Jian-Guo Liu. Spectroscopy system based on a single quantum cascade laser for simultaneous detection of CO and CO2J. Chin. Phys. B, 2016, 25(9): 094210. |
Spectroscopy system based on a single quantum cascade laser for simultaneous detection of CO and CO2
-
Abstract
A quantum cascade laser (QCL) based system for simultaneous detection of CO and CO2 is developed. The QCL can scan over two neighboring CO (2055.40 cm-1) and CO2 (2055.16 cm-1) lines with a single current scan. The wavelength modulation spectroscopy (f = 20 kHz) is utilized to enhance the signal-to-noise ratio. A white cell with an effective optical path length of 74 m is used. The calibration of the sensor is performed and minimum detection limits of 1.3 ppb (1×10-9) for CO and 0.44 ppm (1×10-6) for CO2 are achieved. -
DownLoad: