Cite this article:
Cui Wei-Cheng, Tu Juan, Hwang Joo-Ha, Li Qian, Fan Ting-Bo, Zhang Dong, Chen Jing-Hai, Chen Wei-Zhong. In vivo hyperthermia effect induced by high-intensity pulsed ultrasoundJ. Chin. Phys. B, 2012, 21(7): 074301.
| Cui Wei-Cheng, Tu Juan, Hwang Joo-Ha, Li Qian, Fan Ting-Bo, Zhang Dong, Chen Jing-Hai, Chen Wei-Zhong. In vivo hyperthermia effect induced by high-intensity pulsed ultrasoundJ. Chin. Phys. B, 2012, 21(7): 074301. |
In vivo hyperthermia effect induced by high-intensity pulsed ultrasound
-
Abstract
Hyperthermia effects (39--44 ℃) induced by pulsed high-intensity focused ultrasound (HIFU) have been regarded as a promising therapeutic tool for boosting immune responses or enhancing drug delivery into solid tumor. However, previous studies also reported that the cell death occurs when cells are maintained at 43 ℃ for more than 20 minutes. The aim of this study is to investigate thermal responses inside in vivo rabbit auricular veins exposed to pulsed HIFU (1.17 MHz, 5300 W/cm2, with relatively low-duty ratios (0.2%--4.3%). The results show that: (1) with constant pulse repetition frequency (PRF) (e.g., 1 Hz), the thermal responses inside the vessel will increase with the increasing duty ratio; (2) a temperature elevation to 43 ℃ can be identified at the duty ratio of 4.3%; (3) with constant duty ratios, the change of PRF will not significantly affect the temperature measurement in the vessel; (4) as the duty ratios lower than 4.3%, the presence of microbubbles will not significantly enhance the thermal responses in the vessel, but will facilitate HIFU-induced inertial cavitation events. -
DownLoad: