Precise measurement of a weak radio frequency electric field using a resonant atomic probe*

Project supported by the National Key R&D Program of China (Grant No. 2017YFA0304203), the National Natural Science Foundation of China (Grant Nos. 61475090, 61675123, 61775124, and 11804202), the State Key Program of National Natural Science of China (Grant Nos. 11434007 and 61835007), and Changjiang Scholars and Innovative Research Team in University of Ministry of Education of China (Grant No. IRT 17R70).

Hao Liping1, Xue Yongmei1, Fan Jiabei1, Bai Jingxu1, Jiao Yuechun1, 2, Zhao Jianming1, 2, †
       

Measurements of the lock-in output of the probe transmission (red squares) and signal fields ES (blue circles) by EIT-AT spectra versus the square root of the output power, PRF , of the function generator for (a) signal field frequency ωS ∼ 2π × 2.18 GHz and (b) ωS ∼ 2π × 1.32 GHz. The black line is the calculation of equation ERF=30PRFg/d . The detectable minimum of signal field ESmin = 1.36 ± 0.04 mV/m in (a) and 1.33 ± 0.02 mV/m in (b). The error bars display the standard error of two independent measurements. The inset demonstrates the lock-in output of the probe transmission.