Theory and method of dual-energy x-ray grating phase-contrast imaging
Rong Feng1, 2, †, Gao Yan1, Guo Cui-Juan1, Xu Wei1, Xu Wei1
College of Electronics and Information Engineering, Tianjin Polytechnic University, Tianjin 300387, China
Tianjin Key Laboratory of Optoelectronic Detection Technology and System, Tianjin Polytechnic University, Tianjin 300387, China

 

† Corresponding author. E-mail: shusheng677@163.com

Abstract

The principle of dual-energy x-ray grating phase-contrast imaging (DEPCI) is clarified by using the theory of x-ray interference and Fresnel diffraction. A new method of retrieving phase from the two interferograms is proposed for DEPCI, and its feasibility is verified via simulation. Finally, the proposed method applied to DEPCI experiment demonstrates the effectiveness of the method. This paper lays the theoretical foundation for performance optimization of DEPCI and the further integration of DEPCI and computed tomography.

1. Introduction

Traditional x-ray imaging is based on the varied absorption of x-rays by the different materials in the object of interest. While it is widely used in medical imaging and industrial non-destructive testing, the poor absorption of x-rays by soft tissues in organisms, such as in ligaments and tumors, results in poor absorption contrast. Fortunately, phase contrast imaging can compensate for this deficiency. Absorption and phase contrast of x-rays result from the attenuation of their amplitudes and shifts of their phases, respectively. Upon exposure to hard x-rays (0.01 nm–0.1 nm), the shift of phase is at least three orders of magnitude higher than that caused by the attenuation of amplitude. Even for thin specimens, the presence of light elements can cause an obvious phase contrast. This means that phase-contrast imaging can achieve more detailed imaging of specimens that are unrecognizable when imaged by absorption contrast. Therefore, for objects composed of light elements, detecting a shift of phase is easier and more efficient than detecting an attenuation of the amplitudes.

Since the late twentieth century, various forms of x-ray phase-contrast imaging have gradually been developed. Although the crystal interferometer[1] and free-space propagation techniques[2] yield ideal results for specific problems, none of them has found wide applications in medical or industrial area, which typically require the use of a laboratory x-ray source, a large imaging field of view and good imaging system stability. Grating phase contrast imaging represented by the Talbot–Lau interferometer[3] can solve the above problems. However, its corresponding method of retrieving the phase, called “phase stepping”, requires the multi-stepping of the grating in its single period with the sample existing, which will bring two limitations. The first limitation is that the objects need to be exposed to x-rays at least three times (generally 5–11 times), prolonging the data acquisition time, and causing the ionizing x-ray radiation to be absorbed excessively, which may leads organisms to suffer from genetic disease or even carcinosis. The second limitation is that the grating must be moved in micron-sized steps, thus requiring a high-level mechanical precision. The reduction of radiation doses during x-ray imaging and the number of phase steps have been a focus of recent investigation.

Zhu et al.[4] proposed a method of projecting objects forward and backward without a stepping device for grating in 2010. Their method reduces the exposure time but requires high grating uniformity. Liu et al.[5] proposed a two-step method by combining the analyzer grating with fluorescent screen in 2010. However, a stepping device was still required. In 2017, for free-space propagation techniques, Wang et al.[6] proposed an excellent method, which is expected to be used in zero-step grating phase contrast imaging. In 2010, Kottler et al.[7] applied the dual-energy modality of the traditional x-ray imaging to the Talbot–Lau interferometer, thus enabling the phase contrast imaging to discriminate the information about the specific material composition. With this background, Rong[8] et al. proposed the dual-energy x-ray grating phase contrast imaging (DEPCI) with a dual-energy analyzer grating[9] in 2018. With omitting the stepping device, only the working energy in the DEPCI system needs to be changed once in the imaging experiment. However, accurate theoretical calculations have not yet been made for this method of retrieving the phase. In this paper, the theoretical framework of DEPCI is first elaborated. Furthermore, a new method of retrieving the phase is proposed. The relevant simulation and experimental results demonstrate the effectiveness of this method.

2. Theoretical background

As shown in Fig. 1, the DEPCI system is composed of a normal x-ray source, a source grating G0, a phase grating G1, a dual-energy analyzer grating G2, and a charge coupled device (CCD) detector. The object is placed near the front of G1. All components are centrally coaxial. The G0G1 distance, l, is larger than G1G2 distance, d.

Fig. 1. Model of DEPCI system.

The source grating G0 divides the normal x-rays into several slit x-rays. In particular, x-rays within the same slit are coherent. The Talbot effect occurs when the spatial coherence length meets the condition , where λ, p1, and s denote the wavelength of the x-rays, period of grating G1, and slit width of G0, respectively. The coherent x-rays in each slit produce an interference fringe called the self-image of G1 at the Talbot distance downstream of G1. As shown in Fig. 2, self-images and grating G2 have the same period.[10] In order to improve the transmission efficiency of x-rays, a π phase grating is selected as G1. The m-th order Talbot distance dm can be calculated from[11]

Then, the Lau effect occurs if , where p0 and p2 denote the periods of G0 and G2, respectively. All the interference fringes are staggered by one period and superimposed incoherently. Therefore, the contrast of the total interference fringe is greatly enhanced, even without a highly coherent x-ray source.

Fig. 2. Effect diagram for the dual-energy analyzer grating.

However, the period of a self-image is typically a few microns, which is smaller than the minimum resolution of CCD detector. Therefore, a dual-energy analyzer grating G2 is overlapped with the self-image at an angle to form large-period moirè fringes. The coordinates of the self-image and dual-energy analyzer grating are shown in Fig. 2. A dual-energy analyzer grating is staggered with low-energy and high-energy grid bars. The joints of low- and high-energy grid bars align with the center of bright or dark stripes in the self-image. The low-energy grid bars convert the low-energy x-rays into visible light, and the high-energy grid bars block low-energy x-rays from passing through. The high-energy grid bars convert high-energy x-rays into visible light, and the low-energy grid bars allow high-energy x-rays from passing through directly without being converted to visible light.[8]

The alternating response of the high- and low-energy grid bars of G2 to high- and low-energy x-rays are equivalent to a step in the traditional grating of Talbot–Lau interferometer. Therefore, it is necessary to ensure that the Talbot distance and fringe distribution of the self-images in high- and low-energy x-rays are consistent with each other. According to Eq. (1), if , where and denote the wavelengths of high- and low-energy x-rays, respectively, the third-order Talbot distance of low-energy x-ray imaging is equal to the first-order Talbot distance of high-energy x-ray imaging. Therefore, d can be calculated from

The physical description of object can be expressed by a complex refractive index , where the real part δ and the imaginary part β represent the phase factor and absorption factor, respectively. The formula for the phase factor[12] can be derived from Maxwell’s equations as , where ρ is the mass density, is the classical radius of the electron, is Avogadro’s constant, Z is the atomic number of the element, and is the atomic mass. Therefore, if , the relationship between the phase factors of the high-energy x-rays and the phase factor of the low-energy x-rays is

Assuming that is the phase factor of G1, which has thickness T, the phase change of the x-rays caused by G1 can be calculated from[13]

In DEPCI, G1 is designed to shift the phase of the low-energy x-rays by π. Correspondingly, G1 shifts the phase of the low-energy x-rays by 3π. The self-images of the π and 3π phase grating are consistently distributed.[14]

3. Intensity distribution

As shown in Fig. 3, , , and represent the source-grating, phase-grating, and detector planes, respectively, with being far away from . Furthermore, the size of each slit s is so small that the incident x-rays are approximated as divergent spherical waves.

Fig. 3. Schematic diagram of coherent x-rays propagating in DEPCI system.

The complex wave-front amplitude propagating from the point source a of to is expressed as

where is the transfer function of the phase grating and is a wavevector. After passing through the phase grating, the spherical wave follows the Fresnel–Kirchhoff diffraction. The complex wavefront amplitude at is
where r represents the distance from a point in to a point in . The sample size is much larger than the x-ray wavelength, so it conforms to the paraxial approximation, i.e.,
The approximation is reached by substituting d for r in the denominator of Eq. (6). In the numerator, a change in r will cause the exponential value to change dramatically. Therefore, if r in the numerator takes a second-order approximation as shown as Eq. (7), then equation (6) can be written as
Equation (8) can be reorganized into
with
The Fourier series of the transfer function of G1 is
where is the Fourier coefficient. In DEPCI, , so the magnification . Combining Eq. (9) with Eq. (10) yields

When a sample is placed in the DEPCI system, x-rays undergo amplitude (A(x)) attenuation and phase shift. According to Ref. [15], the transmission function M(x) of the sample is

The phase grating G1 splits the x-rays, producing a shear expressed as . In particular, zeroth-order diffraction results in enhanced background imaging, which reduces the fringe contrast. The phase grating eliminates the zeroth-order diffraction, allowing more than 80% of the intensity to be concentrated on the ± 1st order diffraction.[16] Therefore, in this paper we only consider the ±1st order diffraction and .[17] Combined with Eq. (12), equation (11) can be written as

For hard x-rays in light-element objects, the shear quantity of the amplitude attenuation is quite small, so we can reasonably take the following approximation: . Due to , the intensity distribution at is

where the value of and are calculated from , and ,[18] where is the linear absorption coefficient. , where and are the atomic density and absorption cross section of element i. In the diagnostic range of x-ray energy, the absorption cross section as a function of x-ray energy is shown in Fig. 4. Curves for light elements such as Li, C, N, O are shown. The value of , , , and from the tables published in Ref. [19] are adopted.

Fig. 4. Curves of absorption cross section versus x-ray energy for 4 different light elements.

It should be noted that as the x-ray energy increases, the values of the absorption cross sections tend to be constant and their dimensions are on the order of . This means that the linear absorption coefficient of the low-energy x-ray and the linear absorption coefficient of the high energy x-ray are approximately consistent with each other.

4. Phase recovery method

As shown in Fig. 5, χ represents the relative displacement between the self-image and G2. With alternating high- and low-energy grid bars along the x axis, the normalized intensity of a pixel changes sinusoidally. The duty cycle of G2 is 1:2. When a single pixel contains N periods of self-imaged fringes, the intensity of a single pixel is given by

If we assume the optical coupling efficiency of G2 to be an ideal value, and consider Eq. (14), we can rewrite Eq. (15) as
where is the refraction angle[20] and . In the DEPCI experiment, two interferograms are collected by using the setup as shown in Fig. 1, first at low energy corresponding to , and second at high energy corresponding to . The refraction angle caused by the same object under low- and high-energy x-rays are expressed by and , respectively. The intensity obtained by the detector in low- and high-energy x-ray imagings are denoted by Il and Ih, respectively. According to the expression for θ, we can conclude that . Consequently, the intensities Il and Ih are
Solving the above equations gives
This is similar to the formula mentioned in Ref. [21]. The absorption information of the sample can also be further obtained as follows:

Fig. 5. Model for detecting self-image by using dual-energy analyzer grating.
5. Simulation and experiment
5.1. Simulation of DEPCI system

To demonstrate the feasibility of the proposed method, we simulated the imaging procedure of DEPCI with a PMMA-formvar ball. The ball had an average density , and filled half of the entire field of view. To meet , 20-keV and 60-keV x-rays were selected as sources with low energy and high energy by using .[22] After the appropriate normalization of the two interferograms in intensity, the calculations of the normalized interferograms as shown in Eq. (18) yielded the phase contrast image of the sample, which is shown in Fig. 6(a). Its relevant line profile is shown in Fig. 6(b).

Fig. 6. (a) Phase contrast image, (b) line profile corresponding to red line in panel(a).

As shown in Fig. 6(a), a clear phase contrast image of the PMMA-formvar ball can successfully and clearly be obtained. Furthermore, it can be seen from Fig. 6(b) that the simulation results are consistent with the experimental results from Ref. [3]. The refraction effect is obvious at the edge of the ball, which proves that the method of phase retrieval given by Eq. (18) is applicable to DEPCI.

5.2. Experiments

During the DEPCI experiment, the focal spot size of the conventional tungsten target x-ray source was 1 mm×0.8 mm. An x-ray energy radiation detector was used to calibrate the energy of x-rays. Since the required exposure decreased as the difference between the two energies increased,[23] the wide-spectrum x-rays with the central photon energy of 20 keV and 60 keV were selected as the values of low and high energies, respectively. The tube current was also adjustable from 2 mA to 4 mA. The duty ratios of G0 and G1 were 1:3 and 1:2, respectively. The following parameters were used in the simulation: , , , d=55.7 cm, and l=228.4 cm. The sample was a crayfish, and p2 was coupled with a 2048 × 2048 pixel ( /pixel) CCD camera. The exposure time of each interferogram was 8 s. The phase and absorption contrast images of the sample can be obtained by Eqs. (18) and (19). The effects of the reconstructed images are shown in Figs. 7(a) and 7(c). Furthermore, for the sake of analysing the effect of DEPCI, we performed a four-step phase stepping by using a Talbot–Lau interferometer. The central photon energy of wide spectrum x-rays was 30 keV. The effects of the reconstructed images are shown in Figs. 7(b) and 7(d). Two body parts of the sample marked in Figs. 7(a) and 7(b) were used to fit the data, and the results are shown in Figs. 7(e) and 7(f).

From the extracted information shown in Fig. 7, it can be intuitively seen that the structural characteristics of the sample are successfully demonstrated. The energy of the x-rays used in the Talbot–Lau interferometer was different from that used in DEPCI during the experiments. Illuminating the same object, the x-rays with different energies will produce different phase shifts. In this case, if the trends presented in the two reconstructed phase contrast images are consistent and correspond to the structure of sample, we can determine that the reconstructed images are correct. As can be seen in Figs. 7(e) and 7(f), the trend of the differential phase obtained by the proposed method is almost the same as that obtained by four-step phase stepping.

Fig. 7. (a) Phase contrast image using DEPCI system, (b) phase contrast image using Talbot–Lau interferometer, (c) absorption contrast image using DEPCI system, (d) absorption contrast image using Talbot–Lau interferometer, (e) data fitting for the cross-section of solid line in panels (a) and (b), and (f) data fitting for cross-section of dashed line in panels (a) and (b).

However, the DEPCI has a disadvantage that it cannot be used for dark field imaging. In Ref. [24], equation (16) is written in Fourier series. The 0th and 1st Fourier coefficients constitute the expression of dark field information of object. In the DEPCI, only the ratio of 0th Fourier coefficients can be calculated through the relationship between high and low energies of x-rays, while the 1st Fourier coefficients cannot. Furthermore, the surface contrast in dark field imaging requires sufficient intensity of incident x-rays.[25] From Figs. 7(a) and 7(c), we can see clearly that the intensity obtained by the DEPCI system is weak, compared with the intensity obtained by the Talbot–Lau interferometer. This indicates that the x-ray energy utilization in the DEPCI system was low. Future improvements to the design of the dual-energy analyzer grating and grating fabrication process may break through this limitation.

6. Conclusions and perspectives

In this paper, we present theoretical calculations for DEPCI. Furthermore, by exploiting the intrinsic phase change and characteristics of linear absorption coefficient of low- and high-energy x-rays, a method of retrieving the phase for DEPCI is found. The results of simulation and experiment demonstrate that this method is suitable for DEPCI. The phase-contrast image recovered from the two interferograms collected by the DEPCI without moving any device has a similar effect to the phase-contrast image recovered from the four interferograms collected by the Talbot–Lau interferometer by moving the grating four times. Due to the short exposure time, high stability, and computed tomography (CT) compatibility, the DEPCI may become, we believe, a popular method of obtaining the x-ray grating phase contrast imaging.

Reference
[1] Momose A Takeda T Itai Y Hirano K 1996 Nat. Med. 2 473
[2] Snigirev A Snigireva I Kohn V Kuznetsov S Schelokov A 1995 Rev. Sci. Instrum. 66 5486
[3] Pfeiffer F Weitkamp T Bunk O David C 2006 Nat. Phys. 2 258
[4] Zhu P P Zhang K Wang Z L Liu Y J Liu X S Wu Z Y McDonald S A Marone F Stampanoni M 2010 Proc. Natl. Acad Sci. USA 107 13576
[5] Liu X Guo J C Niu H B 2010 Chin. Phys. 19 070701
[6] Wang F X Wang Y Wei G X Du G H Xue Y L Hu T Li K Deng B Xie H Xiao T Q 2017 Appl. Phys. Lett. 111 174101
[7] Kottler C Revol V Kaufmann R Urban C 2010 J. Appl. Phys. 108 114906
[8] Rong F Xie Y N Tai X F Geng L 2017 Acta Phys. Sin. 66 018701 (in Chinese) https://wulixb.iphy.ac.cn/CN/10.7498/aps.66.018701
[9] Rong F Liang Y Xu W Lei Y H Li J 2018 IEEE Photon. Technol. Lett. 30 1795
[10] Yang J Guo J C Lei Y H Yi M H Chen Li 2017 Chin. Phys. 26 028701
[11] Donath T Chabior M Pfeiffer F Bunk O Reznikova E Mohr J Hempel E Popescu S Hoheisel M Schuster M Baumann J David C 2009 J. Appl. Phys. 106 054703
[12] Lider V V Kovalchuk M V 2013 Crystallogr. Rep. 58 769
[13] Liu X Lei Y H Zhao Z G Guo J C Niu H B 2010 Acta Phys. Sin. 59 6927 (in Chinese) https://doi.org/10.7498/aps.59.6927
[14] Yaroshenko A Bech M Potdevin G Malecki A Biernath T Wolf J Tapfer A Schüttler M Meiser J Kunka D Amberger M Mohr J Pfeiffer F 2014 Opt. Express 22 547
[15] Yan A Wu X Liu H 2008 Opt. Express 16 13330
[16] Weitkamp T Diaz A David C 2005 Opt. Express 13 6296
[17] Yang Q Liu X Guo J C Lei Y H Huang J H Niu H B 2012 Acta Phys. Sin. 61 160702 (in Chinese) https://doi.org/10.7498/aps.61.160702
[18] Momose A 2002 J. Synchrot Radiat. 9 136
[19] Sasaki S 1990 KEK Report 90-16 Tsukuba-shi, Japan National Laboratory for High Energy Physics
[20] Wei C X Wu C Wali F Wei W B Bao Y Luo R H Wang L Liu G Tian Y C 2017 Chin. Phys. 26 108701
[21] Chapman D Thomlinson W Johnston R E Washburn D Pisano E Gmür N Zhong Z Menk R Arfelli F Sayers D 1997 Phys. Med. Biol. 42 2015
[22] Liu H Q Ren Y Q Zhou G Z He Y Xue Y L Xiao T Q 2012 Acta Phys. Sin. 61 078701 (in Chinese) https://doi.org/10.7498/aps.61.078701
[23] Shao Q G Chen J Wali F Bao Y Wang Z L Zhu P P Tian Y C Gao K 2016 Chin. Phys. 25 108702
[24] Pfeiffer F Bech M Bunk O Kraft P Eikenberry F E Brönnimann C Grünzweig C David C 2008 Nat. Mater. 7 134
[25] Daisuke S Sugiyama H Kunisada T Ando M 2006 Appl. Radiat. Isot. 64 868