中国物理B ›› 2010, Vol. 19 ›› Issue (8): 87802-087802.doi: 10.1088/1674-1056/19/8/087802
贾鹏翔1, 包尚联1, 张峰2, 闫镔2
Jia Peng-Xiang(贾鹏翔)a), Zhang Feng(张峰)b), Yan Bin(闫镔)b), and Bao Shang-Lian (包尚联)a)†
摘要: A scattering correction method for a panel detector based cone beam computed tomography system is presented. First, the x-ray spectrum of the system is acquired by using the Monte Carlo simulation method. Secondly, scattered photon distribution is calculated and stored as correction matrixes by using the Monte Carlo simulation method according to scanned objects and computed tomography system specialties. Thirdly, scattered photons are removed from projection data by correction matrixes. A comparison of reconstruction image between before and after scattering correction demonstrates that the scattering correction method is effective for the panel detector based cone beam computed tomography system.
中图分类号: (Computed radiography)